Instructions for use ESTETICA E70 Vision / E80 Vision
4 Operation | 4.17 Using the COMFORTdrive 200 XD/COMFORTbase (optional accessory)
140 / 178
▶
Open the lock
③
and remove the hose.
▶
Remove the hose set from the unit and discard it.
4.17 Using the COMFORTdrive 200 XD/COMFORTbase
(optional accessory)
4.17.1 General use
CAUTION
Failure to comply with the Instructions for Use of the COMFORTdrive
200 XD
Injury to persons or property damage
▶
The use of the COMFORTdrive 200 XD is described in separate Instructions
for Use. Read these instructions before starting up the
COMFORTdrive 200 XD and the COMFORTbase!
The KaVo COMFORTdrive 200 XD is a dental instrument for the high speed
range up to 200,000 rpm. It can be attached only to the KaVo COMFORTbase
coupling.
The hose of the KaVo COMFORTbase is part of the coupling and cannot be re-
moved!
Operation and changing the settings via the control element is identical to the
INTRA LUX motor KL 703.
Also refer to: Settings for the INTRA LUX Motor KL 703 LED and the COM-
FORTdrive
4.17.2 Fitting the motor hose on the dentist's element
▶
Connect the motor hose of the COMFORTbase to the connector for the mo-
tors and pneumatic instruments.
4.17.3 Replace O-rings
CAUTION
Missing or damaged O-rings.
Malfunction and premature failure.
▶
Make sure that all O-rings are on the coupling and are undamaged.
Number of available O-rings: 3
▶
Press the O-ring between your fingers to form a loop.
▶
Push the O-ring to the front, and remove it.
▶
Insert new O-rings (
Mat. no. 1.005.0327
) into the grooves.
Note
The O-ring on the COMFORTbase may only be lubricated with cotton ball wet
with KAVOspray.
See also:
2
Care instructions for the COMFORTbase