Instructions for use ESTETICA E70 Vision / E80 Vision
7 Safety check - Test instructions | 7.2 Instructions for the safety check
7.1.6 Notes on repeat tests
Note
The value determined in these tests must be documented and evaluated to-
gether with the measuring processes. The measured values may not over-
shoot the specified values.
Note
Comparisons with previous measurements must be carried out if the meas-
ured values undershoot the threshold values by more than 10 %. The test in-
tervals should be reduced if a deterioration in values is determined!
7.2 Instructions for the safety check
7.2.1 Preparatory measures on the device
WARNING
Electrical power.
Death or injury from electric shock.
▶
Before servicing, pull the mains plug out of the socket or completely discon-
nect the device from the power to de-energise it!
▶
After conversion, check the electrical safety according to DIN EN 62353
[IEC 62353].
▶
Turn off the main switch before any maintenance work.
The cladding parts defined below must be removed for the safety check:
▶
Release the front cover
④
below and remove.
▶
Loosen screws
⑤
and
⑦
and remove the bracket
⑥
.
Note
For safety reasons, the two side covers are secured with the safety clip and
the associated screws.
This corresponds to the regulation that states that the housing cladding may
only be opened with tools.