Instructions for use ESTETICA E70 Vision / E80 Vision
7 Safety check - Test instructions | 7.2 Instructions for the safety check
158 / 178
ACP
ACP
Note
Additional measuring points ACP X must be taken into consideration in the
presence of accessories.
See also:
2
8 Appendix - Additional measuring sites, Page 167
ACPs on the treatment centre
No ACPs need to be connected to the E70 Vision and E80 Vision treatment
centres during the measurement with PE because all relevant parts have
already been connected with PE in the factory and are included in the test.
ACPs on treatment lamps
No ACPs need to be connected to the operating lights during the measurement
with protective conductor (PE) because all relevant parts have already been
connected to the protective conductor (PE) at the factory and are included in
the test.
Measuring the protective conductor resistance
Limit
< 0,3 Ω (maximum value!)
Note
The integrity of the power supply cable, in particular the protective earth wire
of the power cable must be ensured. As this is a fixed installation, the evalu-
ation can be conducted by means of a visual inspection. If damage is determ-
ined, the further procedure to be taken is specified in the general instruc-
tions.
Note
In this measurement the resistance of the protective earth connection of the
supply network can be taken into consideration.
Note
If applicable: all removable supply connection lines, which are kept handy for
possible use, should be taken into consideration and the respective PE meas-
ured.