Instructions for use ESTETICA E70 Vision / E80 Vision
4 Operation | 4.10 Using functions through the touchscreen
70 / 178
ð
This causes the settings menu of the multifunctional handpiece to be dis-
played.
Adjusting the air/water heating
▶
Tap the corresponding key to set "Heating air/water" as desired.
Symbol
Function
Air/water heating "On"
Air/water heating "Off"
▶
After you set a single value or all values, tap the "Save" key to save the val-
ues.
Operation of the treatment light KaVoLUX 540 LED U
WARNING
Unintentional activation of the KaVo KEY Laser III and KEY Laser 3+.
Simultaneous application of the operating light KaVoLUX 540 LED and the
KaVo KEY Laser III or KEY Laser 3+ can lead to the unintentional activation of
the KaVo KEY Laser III and KEY Laser 3+.
▶
When using the KaVo KEY Laser III or the KEY Laser 3+, switch the operat-
ing light to laser mode.
▶
Or switch off the operating light, do not use the KaVo KEY Laser III or
KEY Laser 3+ and the KaVoLUX 540 LED operating light simultaneously.
WARNING
Wrong handling.
Reversible blinding (temporary sight impairment).
▶
Do not direct the light field at patients, users or/and third parties while you
switch on the light.
▶
Do not direct the light field at the patient's eyes when you move the light
head.
▶
Maintain a clearance of approx. 700 mm between the operating light and
the mouth of the patient.