Instructions for use EXPERTtorque E680 - 1.006.4700EXPERTtorque Mini E677 - 1.006.0100
7 Processing steps in accordance with ISO 17664 | 7.5 Packaging
20 / 26
▶
Remove the service coupling chuck from the side hatch of the QUATTRO-
care PLUS and attach it to coupling service station four, on the far right. A
MULTIflex adaptor must be mounted there.
▶
Press the handpiece together with the guide bush of the chuck to be ser-
viced against the tip of the service coupling.
▶
Press the button marked with the chuck service symbol.
Note
Close the chuck service mode.
Option 1: Place the dental handpieces in the QUATTROcare PLUS 2124 A,
close the front door and start the service procedure.
Option 2: After three minutes with no service procedure running, the device
automatically switches back to normal service mode.
See also:
2
Servicing with KaVo QUATTROcare PLUS
7.5 Packaging
Note
The sterilization bag must be large enough for the handpiece to fit without
stretching the bag.
The quality and use of the sterilization packaging must comply with applica-
ble standards and be suitable for the sterilization procedure!
▶
Seal each medical device individually in a sterilization item package.
7.6 Sterilization
Sterilization in a steam sterilizer (autoclave) in
accordance with ISO 17665-1
CAUTION
Improper service and care.
Risk of injury.
▶
Perform regular proper care and servicing.
NOTICE
Contact corrosion due to moisture.
Damage to the product.
▶
Remove the product from the steam sterilization immediately after the
sterilization cycle.
The KaVo medical device has a maximum temperature resistance of up to
138
℃
(280.4 °F).
Sterilization parameters:
Select a suitable procedure (depending on the available autoclave) from the fol-
lowing sterilization processes: