Instructions for use EXPERTtorque E680 - 1.006.4700EXPERTtorque Mini E677 - 1.006.0100
1 User instructions
4 / 26
1 User instructions
Dear User,
congratulations on purchasing this KaVo quality product. By following the in-
structions below you will be able to work smoothly, economically and safely.
© Copyright by KaVo Dental GmbH
Please direct all questions regarding the product, service and maintenance to
the
KaVo Technical Service:
Toll-free: 1-888-ASK-KAVO (888-275-5286)
Email: customerservice@kavokerr.com
Please refer to the serial number of the product in all inquiries!
For repairs, please contact KaVo Repair Service.
For scheduling or if you have any questions, please contact:
KaVo Repair Service
KaVo Dental Technologies, LLC
11727 Fruehauf Drive
Charlotte, NC 28273 USA
Toll-free Direct Customer Service: 1-888-ASK-KAVO (888-275-5286)
Email: techservice@kavokerr.com
www.kavousa.com
General marks and symbols
Refer to the chapter on Safety/Warning symbol
Important information for users and service technicians
Action request
Sterilization parameters
▪ Sterilizer with triple pre-vacuum:
- at least 3 minutes at 135°C (275 °F)
- Drying time: 16 min.
▪ Sterilizer using the gravity method:
- at least 10 minutes at 135°C (275 °F)
- Drying time: 30 min.
▪ Sterilizer using the gravity method:
- at least 60 minutes at 121°C (250 °F)
- Drying time: 15 min.
Suitable for disinfection in a washer disinfector
Information on the package labeling
Catalogue number
Summary of Contents for EXPERTtorque E680