Instructions for use EXPERTtorque E680 - 1.006.4700EXPERTtorque Mini E677 - 1.006.0100
3 Description of the product | 3.2 Technical Specifications
Proper Use:
In accordance with these regulations, this medical device may only be used by
a properly trained user and for the application described herein. You need to
comply with the following:
▪ the applicable health and safety regulations
▪ the applicable accident prevention regulations
▪ these Instructions for use
In accordance with these regulations, the user is required to:
▪ only use equipment that is operating properly
▪ adhere to the specified intended use
▪ protect himself or herself, the patient and third parties from danger, and
▪ avoid contamination from the product
3.2 Technical Specifications
NSK PTL-CL-LED3
Drive pressure (bar)
2.0 - 3.0
Drive pressure
recommended (bar)
2.5
Return air pressure (bar) < 0.5
Spray water pressure
(bar)
0.8 - 2.0
Spray air pressure (bar)
1.0 - 2.5
NSK PTL-CL-LED3
Air consumption (Nl/min) 39 - 51
Idling speed (rpm) EX-
PERTtorque / EXPERT-
torque Mini
340,000 - 420,000 / 400,000- 480,000
Recommended
contact force (N)
2 - 3
Note
NSK PTL - CL- LED 3 is a product name of NSK NAKANISHI INC., Japan.
3.3 Transportation and storage conditions
▶
Do not store in a refrigerated environment.
Temperature: -20 °C to +70 °C (-4 °F to +158 °F)
Relative humidity: 5% RH to 95% RH absence of condensation
Air pressure: 700 hPa to 1060 hPa (10 psi to 15 psi)
Protect from moisture (Keep dry)