7 Processing steps in accordance with ISO 17664 | 7.3 Automated processing
7.2.2 Manual external disinfection
WARNING
Incomplete disinfection.
Infection hazard.
▶
Only use disinfection procedures that are verified to be bactericidal, fungici-
dal and virucidal.
▶
If the disinfectants used do not meet these requirements, the process must
be concluded by disinfection of the unit(s) without packaging using a steril-
izer.
NOTICE
Never disinfect the handpiece with chlorine-containing products.
Functional damage and property damage.
▶
Only disinfect in a washer disinfector or manually.
KaVo recommends the following products based on material compatibility. The
microbiological efficacy must be ensured by the disinfectant manufacturer and
proven by an expert opinion.
▪ CaviCide made by Metrex
Consumables required:
▪ Cloths for wiping the medical device.
▶
Spray the disinfectant on a cloth, then wipe down the medical device and
allow the disinfectant to act in accordance with the instructions of the
disinfectant manufacturer.
▶
Comply with the instructions for use of the disinfectant.
7.2.3 Manual drying
▶
Clean the outside and inside with compressed air until no drops of water are
visible.
▶
Immediately after drying, lubricate the KaVo medical device with care
agents from the KaVo care system.
See also:
2
7.4 Care products and systems - Servicing, Page 18
7.3 Automated processing
WARNING
Incomplete disinfection.
Infection hazard.
▶
Only use disinfection procedures that are verified to be bactericidal, fungici-
dal and virucidal.
▶
If the disinfectants used do not meet these requirements, the process must
be concluded by disinfection of the unit(s) without packaging using a steril-
izer.
17 / 26