Instructions for use EXPERTtorque E680 - 1.006.4700EXPERTtorque Mini E677 - 1.006.0100
7 Processing steps in accordance with ISO 17664 | 7.7 Storage
▪ Sterilizer with triple pre-vacuum:
- at least 3 minutes at 135°C (275 °F)
- Drying time: 16 min.
▪ Sterilizer using the gravity method:
- at least 10 minutes at 135°C (275 °F)
- Drying time: 30 min.
▪ Sterilizer using the gravity method:
- at least 60 minutes at 121°C (250 °F)
- Drying time: 15 min.
▶
Remove contra-angle handpieces and turbines immediately after the com-
pletion of the sterilization cycle from the sterilizer.
▶
Use in accordance with the manufacturer's Instructions for Use.
7.7 Storage
Processed products must be stored, protected from bacteria, to the extent pos-
sible, and dust, in a dry, dark, cool room.
Note
Comply with the expiration date of the sterilized items.
21 / 26