3 Pre-installation
FOCUS
7
3 Pre-installation
3.1 Pre-installation checklist
Radiation protection:
Follow all local regulations for radiation shielding of
the X-ray room.
During the exposure the operator must follow
applicable radiation shielding requirements and
stay at least 7 feet (2 meters) from the radiation
source.
Mounting requirements:
The wall and mounting hardware must sustain a
pull of 350 lbs./1500N, from each wall bracket bolt.
When replacing an old X-ray unit using a wide wall
mount plate, observe the space that is covered by
the wall mount plate to minimize painting where
applicable.
Mounting on a treatment unit is allowed only when
the manufacturer of the treatment unit has tested
the mechanical stability and allow the integration.
For KaVo treatment unit integration refer the
compatibility table of the accessories available.
For OpenMount installation, the mounting shall
contain a bearing designed for a shaft of diameter
of 32 mm (h7) and a length of 105 mm when
mounting on a 3rd party adapter, to ensure that
the maximum load ratings of the adapter are not
exceeded by the FOCUS OpenMount unit. The
installation must provide tilt adjustment to level
and balance the arm. The installation must not
allow rotating the X-ray unit more than 360
degrees to prevent twisting of the cables.
Electrical requirements:
Mains voltage:
Verify the mains voltage (3 wire: single-phase, 115
V or 230 V with protective ground).
Fuses:
Summary of Contents for FOCUS 215720
Page 2: ......
Page 4: ......
Page 6: ...ii...
Page 12: ...2 Unit description 6 FOCUS Exposure time control...
Page 55: ...6 Dimensions FOCUS 49 6 Dimensions Wide wall mount plate measures in mm...
Page 56: ...6 Dimensions 50 FOCUS Narrow wall mount plate measures in mm...
Page 57: ...6 Dimensions FOCUS 51 Remote control frame measures in mm...
Page 58: ...6 Dimensions 52 FOCUS Wall mount installation dimensions...
Page 59: ...6 Dimensions FOCUS 53...
Page 60: ...6 Dimensions 54 FOCUS...
Page 61: ...6 Dimensions FOCUS 55...
Page 62: ...6 Dimensions 56 FOCUS OpenMount installation dimensions...