5 Finalizing the installation
48
FOCUS
5.3 System checks before use
1.
Check that the device does not move and that
the arm is not drifting.
2.
Check the local regulations on radiation safety.
Carry out the required radiation measurements
and inform the local authorities.
3.
Move the tube head to up and down to check
that the movements are smooth to both
directions and that the tube head stays in
position.
4.
Check Protective Earth (PE) grounding
resistance according to the IEC 62353 (Medical
electrical equipment - Recurrent test and test
after repair of medical electrical equipment).
5.
Check visually that the exposure indicator light
illuminates for the duration of the exposure.
6.
Confirm that the audible indicator beeps for the
duration of the exposure.
7.
Check all functions of the remote control.