8 User's statement
FOCUS
43
8 User's statement
Instructions for the use of the unit and precautionary
statements are part of the unit
User's Manual.
RADIATION LEAKAGE TECHNIQUE FACTORS
The maximum-rated peak tube potential is 70 kVp with the
maximum rated continuous tube current of 1.5 mA.
BEAM LIMITING DEVICE / TUBE HOUSING ASSEMBLY
COMPATIBILITY
The tube housing assembly THA-I is compatible with the
beam limiting device.
EQUIPMENT STATEMENT FOR TUBE HOUSING
ASSEMBLY
Maximum operating voltage is 70 kVp. Nominal focal spot
is 0.7 mm.
X-ray tube: D-0711SB or D-0711S. For additional
information please refer to the tube specification sheets.
Art.No
Part number
Cones
Program
Mode Pr2
Cone type
ID
0.805.0057
50410-img
Long cone, round
SSD 305 mm/12”, Ø60 mm
3
0.805.0058
50420-img
Long cone, rectangular
SSD 305 mm/12”, 35x45 mm
4
0.805.0059
50540-img
Short cone, round
SSD 229 mm/9”, Ø60 mm
1
0.805.0060
50550-img
Short cone, rectangular
SSD 229 mm/9”, 35x45 mm
2
0.805.0061
50551-img
Short cone, rectangular, small
SSD 229 mm/9”, 28x36 mm
5
0.805.0062
50720-img
Long cone, full metal, rectangular
SSD 305 mm/12”, 35 x 45 mm
4
0.805.4235
50750-img
Short cone, full metal, rectangular
SSD 229 mm/9”, 35 x 45 mm
6