background image

CAUTION

Injury hazard from cutters or grinders.

Infections or cuts.

▶ Wear gloves or fingerstalls.

CAUTION

Hazard from defective chucking system.

The cutter or grinder can fall out and cause damage.

▶ Pull on the cutter or grinder to check if the chucking system is okay

and the cutter or grinder is securely held. Fur checking, inserting and

removing, use gloves or a fingerstall to prevent an injury or infection.

Operation

30

Summary of Contents for INTRA 3610 N3

Page 1: ...Instructions for use INTRA surgical handpiece 3610 N3 REF 0 524 5620...

Page 2: ...Sales KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Tel 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Page 3: ...onditions 8 2 Safety 10 2 1 Description of safety instructions 10 2 2 Purpose Proper use 12 2 3 Safety instructions 14 3 Product description 18 3 1 Technical data 18 3 2 Transportation and storage con...

Page 4: ...2 5 5 Conversion for short handpiece cutters or grinders 33 6 Preparation methods according to ISO 17664 35 6 1 Preparations at the site of use 35 6 2 Disassemble the medical device 36 6 3 Assembling...

Page 5: ...ion internal 43 6 5 3 Disinfection Machine disinfection external and internal 44 6 6 Drying 44 6 7 Care products and systems Servicing 45 6 7 1 Care products and systems Servicing Care with KaVo Spray...

Page 6: ...new high quality product Following the instructions below will allow you to work smoothly economically and safely Copyright by KaVo Dental GmbH Symbols See the section Safety Warning Symbols Importan...

Page 7: ...mmunaut Europ enne A product with this mark meets the requirements of the applicable EC directive Action request Target group This document is intended for dentists and their assistants The section on...

Page 8: ...Other claims of any nature whatsoever in particular with respect to compensation are excluded In the event of default gross negligence or intent this shall only apply in the absence of mandatory lega...

Page 9: ...at can be attributed to interventions on or changes made to the product by the customer or a third party not authorised by KaVo are excluded from the warranty Service warranty claims will only be acce...

Page 10: ...ructions Warning symbol Structure DANGER The introduction describes the type and source of the danger This section describes the possible consequences of misuse The optional step contains necessary me...

Page 11: ...e three levels of danger will help avert property damage and injury CAUTION CAUTION indicates a hazardous situation that can lead to property damage or minor to moderate injury WARNING WARNING indicat...

Page 12: ...roper use This medical device is only for dental treatment by a dental professional The product may not be changed or used for any other purpose since this may be hazard ous The medical device is inte...

Page 13: ...pplicable health and safety regulations the applicable accident prevention regulations these instructions for use According to these regulations the user is required to only use properly operating equ...

Page 14: ...CAUTION Premature wear and malfunctioning from improper storage during long periods of nonuse Shortened product life The instrument must be cleaned serviced and stored dry if it has not been used for...

Page 15: ...cutter or grinder cannot be firmly held Stop work and seek service support CAUTION Hazard from improperly putting away instruments Injury and infection caused by chucked cutters or grinders Damage to...

Page 16: ...cians specially trained by KaVo To ensure proper function the medical device must be set up according to the methods described in the KaVo instructions for use and the care prod ucts and methods descr...

Page 17: ...Service may only be provided by repair shops that have undergone training by KaVo and that use original KaVo replacement parts Safety 17...

Page 18: ...description INTRA surgical handpiece 3610 N3 Mat no 0 524 5620 3 1 Technical data Drive speed max 40 000 rpm 1 Identification 1 blue ring Transmission 1 1 Maximum speed max 40 000 rpm 1 Product descr...

Page 19: ...use Page 22 Handpiece cutters or grinders can be inserted After conversion short handpiece cutters or grinders can be inserted The handpiece can be mounted on all INTRAmatic Lux motors and motors with...

Page 20: ...hazardous after it has been stored in an excessively cold location The medical device can malfunction Prior to start up very cold products must be heated to a temperature of 20 C 68 F to 25 C 77 F Tem...

Page 21: ...Air pressure 700 hPa 10 psi to 1060 hPa 15 psi Protect from moisture Product description 21...

Page 22: ...4 First use WARNING Hazard from unsterile products Infection hazard to the dentist and patient Sterilize the medical device before initial use and after each treat ment First use 22...

Page 23: ...in emphysema Avoid the insufflation of irrigation in open wounds in surgical area Turn off the spray air and spray water supply on the treatment unit Cool the cutter or grinder using an external suppl...

Page 24: ...Do not use a different coolant First use 24...

Page 25: ...ith the drive motor Lock the handpiece Only start the hand piece when the chuck is closed CAUTION Removing and attaching the handpiece while the device is rotating Damage to the catch Never attach or...

Page 26: ...e that is not properly locked in place can release from the supply hose Carefully pull on it before each treatment to ensure that the medical device is securely locked on the supply hose Place the med...

Page 27: ...ure that the medical device is securely affixed to the coupling 5 2 Remove the medical device Unlock the medical device from the motor coupling and remove it or pull it off by twisting it slightly on...

Page 28: ...tters or diamond grinders that correspond to ISO 1797 1 type 3 are made of steel or hard metal and meet the following criteria shaft diameter 2 334 to 2 350 mm without drill stop Overall length max 70...

Page 29: ...tructions for use and use the cutter or grinder properly Only use cutters or grinders that do not deviate from the indicated data CAUTION Injury from using worn cutters or grinders Cutters or grinders...

Page 30: ...from defective chucking system The cutter or grinder can fall out and cause damage Pull on the cutter or grinder to check if the chucking system is okay and the cutter or grinder is securely held Fur...

Page 31: ...all the way in the direction of the arrow and insert the handpiece cutter or grinder into the chuck Turn the grip sleeve back into its initial position Check that the cutter or grinder is seated by pu...

Page 32: ...ing system Do not touch rotating cutters or grinders Never press the pushbutton while the cutter or grinder is rotating Do not touch soft tissue with the head tip that can heat and burn the soft tissu...

Page 33: ...der Turn the grip sleeve back into its initial position 5 5 Conversion for short handpiece cutters or grinders Note The handpiece must be converted when short cutters or grinders are used Open the han...

Page 34: ...Remove the drill stop using the accompanying hook Operation 34...

Page 35: ...fection hazard exists from contaminated medical devices Observe suitable personal protective measures Note After each operation immediately disassemble the medical device and clean it thoroughly to pr...

Page 36: ...the like Prepare the medical device directly before treatment Remove cutters or burs from the medical device 6 2 Disassemble the medical device Remove cutters or burs from the medical device Grip the...

Page 37: ...nctions from cleaning in the ultrasonic unit Defects in the product Only clean in a thermodisinfector or manually 6 4 1 Cleaning Manual cleaning external Required accessories Tap water 30 C 5 C 86 F 1...

Page 38: ...KaVo DRYspray Cover the medical device with the KaVo CLEANpac bag and place it on the corresponding care adapter Press the spray button three times for 2 seconds each time Remove the medical device fr...

Page 39: ...Luxem bourg Belgium Netherlands United Kingdom Denmark Sweden Fin land and Norway In other countries only automated interior cleaning with thermodisinfectors in accordance with ISO 15883 1 6 4 4 Clea...

Page 40: ...ly refers to material compatibility with KaVo products The program settings and cleansers and disinfectants that must be used can be found in the instructions for use of the thermodisinfector Directly...

Page 41: ...ion CAUTION Malfunctioning from using a disinfectant bath or chlorine containing dis infectant Defects in the product Only disinfect in a thermodisinfector or manually Preparation methods according to...

Page 42: ...fection external KaVo recommends the following products based on material compatibility The microbiological efficacy must be ensured by the disinfectant manufac turer CaviCide by Metrex Preparation me...

Page 43: ...the disinfectant manufacturer disinfectant manufacturer Note Observe the instruction for use for the disinfectant 6 5 2 Disinfection Manual disinfection internal For effective preparation the interior...

Page 44: ...eansers and disinfectants that need to be used can be found in the instructions for use for the thermodisinfector Directly after automated cleaning treat the medical device with the care products and...

Page 45: ...program of the thermo disinfector 6 7 Care products and systems Servicing CAUTION Sharp cutters or grinders in the medical device Injury hazard from sharp and or pointed cutters or grinders Remove th...

Page 46: ...ened product life Regularly perform proper care Note KaVo only guarantees that its products will function properly when the care products are used that are listed as accessories since they were tested...

Page 47: ...noon and in the evening after hours after each time the machine is cleaned and before each sterilisation Remove the cutter or grinder and close the chuck Cover the product with the Cleanpac bag Place...

Page 48: ...care KaVo recommends cleaning and maintaining the chucking system every evening Remove the cutter or grinder place the spray nipple tip in the opening and spray Preparation methods according to ISO 17...

Page 49: ...re with KaVo QUAT TROcare Cleaning and care unit with expansion pressure for thorough cleaning and care KaVo recommends servicing the project twice daily at noon and in the evening after hours after e...

Page 50: ...tter or grinder Servicing the product Chuck care KaVo recommends cleaning and maintaining the chucking system once a week See also Instructions for use KaVo QUATTROcare Preparation methods according t...

Page 51: ...Remove the cutter or grinder place the spray nipple tip in the opening and spray Subsequently treat with the care products and systems listed below Preparation methods according to ISO 17664 51...

Page 52: ...t taut The quality and use of the sterilized product packaging must satisfy ap plicable standards and be suitable for the sterilization procedure Individually weld the medical device in the sterilized...

Page 53: ...oclave EN 13060 ISO 17665 1 CAUTION Premature weary and malfunctions from improper servicing and care Reduced production time Before each sterilization cycle treat the medical device with KaVo care pr...

Page 54: ...ion cycle immediately remove the product from the Steam sterilizer The medical device has a maximum temperature resistance up to 138 280 4 F Autoclave with prevacuum for at least 4 minutes at 134 C 1...

Page 55: ...tion method for at least 60 minutes at 121 C 1 C 250 F 1 8 F Drying time 30 min Follow the manufacturer s instructions for use 6 10 Storage Prepared products should be stored protected germ free from...

Page 56: ...ry Mat no Instrument stand 2151 0 411 9501 Cleanpac 10 units 0 411 9691 Cellulose pad 100 units 0 411 9862 Nozzle needle 0 410 0931 Spray hose sterilisable 0 065 5188 Drill stop 0 524 0892 Hook 0 410...

Page 57: ...Only for the USA Material summary Mat no KAVOspray America 2113 A 0 411 9660 QUATTROcare plus Spray America 2141 P 1 005 4524 Tools 57...

Page 58: ......

Page 59: ......

Page 60: ...1 007 2724 kb 20090526 03 en US...

Reviews: