Instructions for use Primus 1058 Life
9 Troubleshooting
Malfunction
Cause
Remedy
No cold light on the
handpieces.
The high-pressure lamp
or Multi LED on the
handpiece is defective.
▶
Replace the high-pressure lamp or Multi LED.
See also:
2
Instructions for Use of the handpiece
No heating function on
the multifunctional
handpiece.
Spray heating not
preselected.
▶
Pre-select spray heating.
No cold light on the
multifunctional hand-
piece.
The heating function is
preselected.
▶
Deselect the heating function (and select it
again, if applicable).
No spray in the instru-
ments.
No spray preselected.
▶
Preselect spray.
Close the ring for con-
trolling the spray on
the instruments.
▶
Open the ring for controlling the spray on the
instruments.
Spray at the instru-
ments is insufficient.
The spray nozzles are
dirty/clogged.
▶
Clean the spray nozzles according to the ac-
companying instrument operating instructions.
Leaks in instruments.
O-rings at MULTIflex or
motor coupling, grip-
ping sleeve or cannula
of the triple-function
handpiece are dam-
aged.
▶
Replace O-rings.
The suction hoses do
not have any suction.
Sliders on the conical
sections are closed.
▶
Open the sliders.
Sieves in suction con-
nector are clogged.
▶
Replace sieves.
Chair kick plate is being
actuated.
▶
Release the chair kick plate.
Water in the return air
filter.
O-rings of the MUL-
TIflex coupling are
damaged.
▶
Replace all O-rings of the MULTIflex coupling.
A melody sounds.
The amalgam separator
CAS1 is almost full
(95%).
▶
Exchange the amalgam container.
The CAS1 amalgam
separator is defective.
▶
Also refer to:
Instructions for use for the CAS
1 or
▶
Call a Service technician.
Ten beeps are issued.
The Oxygenal container
is too full.
▶
Stop filling the Oxygenal container.
115 / 126