Instructions for use KaVoLUX 540 LED U / T
4 Assembly | 4.3 Assembly of the KaVoLUX 540 LED U
▶
To check the CAN resistance, measure between connections X86.A1 and
X86.A2.
ð
The resistance value must be 60 Ohm.
Note
No adjustments need to be made to the CAN circuit patient chair The follow-
ing table will provide full information.
CAN circuit patient chair
Chair con-
trol
Motor con-
trol lift
Backrest
chair control
Lengthwise
extension
motor con-
trol
Motorised
headrest
control
Switch
S1
SW200
SW200
SW200
S2
E80 all SN
with mot.
headrest
off
on
off
off
on
without mot.
headrest
on
on
off
off
n. a.*
E70 all SN
with mot.
headrest
off
on
n. a.*
n. a.*
on
without mot.
headrest
on
on
n. a.*
n. a.*
n. a.*
* not available / nicht vorhanden
Final tasks
▶
Mount caps on the swinging and spring arm.
▶
Run function test.
▶
Perform the safety check according to the treatment centre.
▶
Mount the device covers.
4.3.3 ESTETICA E70/E80 Vision
WARNING
Electrical power.
Death or injury from electric shock.
▶
Before servicing, pull the mains plug out of the socket or completely discon-
nect the device from the power to de-energise it!
▶
After conversion, check the electrical safety according to DIN EN 62353
[IEC 62353].
37 / 110