8 Special procedures
148
the host PC notice the change and update its routing
tables.
8.10.2
Problem: Automatic software update
fails
Indication
: Failure message from the automatic updater.
Possible reasons and solutions: Automatic software
update may fail for example due to connectivity problems.
Problems may also occur when updating from early
embedded software versions. In case of failure, try to
restart the automatic update. If it fails repeatedly, revert to
manual update.
8.10.3
Problem: Manual update fails
Indication
: Messages printed to the s2terminal indicate
that the update did not succeed.
Possible reasons and solutions
:
1.
Faulty internal or external Ethernet cables. Retry the
update. Don't power off the unit since this will typically
leave it in a non-responsive state! Try to replace cables
without powering off the unit.
2.
Interference by another s2terminal or driver in the host
PC. Make sure any imaging application is closed down.
Close all other s2terminals than the one performing
software update.
Faulty circuit board or insufficient power supply. Measure
the voltage from the circuit board's power connector.
Check that it matches the board's nominal voltage
(typically 5V). Replace the board if faulty (if no other
reason for found for the update to fail).
8.10.4
Problem: S2terminal is unable to find
binary files
Indication
: When trying to program, there is an error
message in s2terminal:
Unable to locate binary file "OP3DPRO_fw_r2_00.srec"
or ".\OP3DPRO_r2_00\OP3DPRO_fw_r2_00.srec"
Possible reasons and solutions
:
Summary of Contents for ORTHOPANTOMOGRAPH OP 3D Pro
Page 2: ......
Page 4: ......
Page 8: ...iv rev...
Page 18: ...1 Introduction 10...
Page 27: ...2 Unit description 19 2 6 Unit movements Panoramic unit movements R C J D...
Page 29: ...2 Unit description 21...
Page 30: ...2 Unit description 22...
Page 38: ...4 PC requirements 30...
Page 54: ...5 Installing the unit 46 4 Fasten the covers...
Page 67: ...5 Installing the unit 59 C C D D Top cover of the column E E...
Page 104: ...7 Cephalometric calibration and alignment 96...
Page 111: ...7 Cephalometric calibration and alignment 103 3 Tighten the locking nut...
Page 117: ...7 Cephalometric calibration and alignment 109...
Page 128: ...8 Special procedures 120 6 Remove the arm from the column 4 nuts 7 Disassemble the arm...
Page 132: ...8 Special procedures 124 right handed assembly Prism assembly for the left handed cephalostat...
Page 158: ...8 Special procedures 150...
Page 165: ...9 OP 3D Pro Pre sales check list 157 9 11 Notes comments...
Page 166: ...9 OP 3D Pro Pre sales check list 158 9 12 Dimensions 463...
Page 167: ...9 OP 3D Pro Pre sales check list 159 9 13 Dimensions with cephalostat...