6 Panoramic and 3D calibrations and adjustments
74
6.5.6
Panoramic
geometry calibration
1. Select the
Pan Geom
program.
2. Press
Patient Positioning
.
3. Install the chin rest and
double cone calibration
tool
.
4. Take an exposure.
5. Repeat the calibration until calibration result “passed” is
achieved.
Chin rest and double cone calibration tool.