6 Panoramic and 3D calibrations and adjustments
79
2. Select the
3D lasers
program.
3. Press
Patient Positioning
.
NOTICE! Lasers can be turned on by pushing the light
button. Lasers stay on approx. 30 seconds.
4. Align the 3D lasers of MFOV (Maxio) by using the MFOV
3D lasers alignment tool (L-shaped metal plate). This tool
is fixed to the phantom holder by using the 3D geometry
calibration phantom.
5. Locations of the lasers and tools:
3D lasers alignment tool