9 OP 3D Pro Pre-sales check list
152
9.3 Mechanical Specifications
9.4 Electrical Specifications
Installation space:
Pan/3D:
Minimum installation space
Depth: 1700 mm
Width: 1500 mm
Height: 2410 mm at minimum
(max height adjustable from
2110 to 2410)
Cephalostat:
Minimum installation space
Depth: 1700 mm
Width: 2500 mm
Height: 2410 mm at minimum
(max height adjustable from
2110 to 2410)
End User/Dealer
Fixing hardware
Required fixing hardware de-
pends on the wall and floor
material and it is not delivered
with the unit. The fixing hard-
ware, as well as wall and floor
materials must endure a pull-
out force of 5000 N. Wall ma-
terial should be suitable for fix-
ing the unit. If the wall is made
of weak material, you m y have
to use a reinforcing plate on
the rear side of the wall to hold
the fixing hardware.The per-
son installing the unit is re-
sponsible for fixing hardware.
Check that it is possible to drill
holes in the floor without dam-
age to any electrical or water
pipes etc.
Dealer
Weight:
The fully assembled unit with
cephalostat weighs 245 kg/
540 lbs (212 kg / 467 lbs with-
out cephalostat). The area of
the column base plate is 0,07
m
2
. Make sure that the floor
where the unit is to be installed
can support this weight.
End User/Dealer
Power network
100-240VAC, 50/60Hz
(10A@230VAC,
15A@110VAC) dedicated
power supply, max. 0.2
Ω
line
impedance. Separate outlets
for the unit and Workstation
PC.
End User/Dealer
Summary of Contents for ORTHOPANTOMOGRAPH OP 3D Pro
Page 2: ......
Page 4: ......
Page 8: ...iv rev...
Page 18: ...1 Introduction 10...
Page 27: ...2 Unit description 19 2 6 Unit movements Panoramic unit movements R C J D...
Page 29: ...2 Unit description 21...
Page 30: ...2 Unit description 22...
Page 38: ...4 PC requirements 30...
Page 54: ...5 Installing the unit 46 4 Fasten the covers...
Page 67: ...5 Installing the unit 59 C C D D Top cover of the column E E...
Page 104: ...7 Cephalometric calibration and alignment 96...
Page 111: ...7 Cephalometric calibration and alignment 103 3 Tighten the locking nut...
Page 117: ...7 Cephalometric calibration and alignment 109...
Page 128: ...8 Special procedures 120 6 Remove the arm from the column 4 nuts 7 Disassemble the arm...
Page 132: ...8 Special procedures 124 right handed assembly Prism assembly for the left handed cephalostat...
Page 158: ...8 Special procedures 150...
Page 165: ...9 OP 3D Pro Pre sales check list 157 9 11 Notes comments...
Page 166: ...9 OP 3D Pro Pre sales check list 158 9 12 Dimensions 463...
Page 167: ...9 OP 3D Pro Pre sales check list 159 9 13 Dimensions with cephalostat...