3 Device upgrade
14.
Connect the 3D sensor power cable.
15.
Guide the cables neatly against the sensor holder frame so that the sensor covers are easy to
install.
3.14 Install CEPH secondary collimator
Install the CEPH secondary collimator to the PAN/3D tubehead and connect all remaining cables to
i2000 circuit board.
1.
Remove the 2 indicated screws and loosen the vertical alignment screw (circled) on the secondary
collimator plate.
2.
Remove the collimator plate from the holder.
ORTHOPANTOMOGRAPH
™
OP 3D
35