6 Reprocessing steps in accordance with ISO 17664 | 6.4 Automated reprocessing
18 / 24
6.4 Automated reprocessing
WARNING
Incomplete disinfection.
Infection hazard.
▶
Only use disinfection procedures that are verified to be bactericidal, fungi-
cidal and virucidal.
▶
If the disinfectants used do not meet these requirements, the process must
be concluded by disinfection of the unit(s) without packaging using a steam
steriliser.
WARNING
Sharp tool in the medical device.
Injury hazard from sharp and/or pointed tool.
▶
Remove the tool.
NOTICE
Never disinfect the medical device with chloride-containing products.
Malfunction and material damage.
▶
Only disinfect in a washer disinfector or manually.
NOTICE
Never reprocess this medical device in an ultrasonic device.
Malfunction and material damage.
▶
Clean manually or in a washer disinfector only.
6.4.1 Preparation for automated internal and external
cleaning as well as internal and external disinfection
Note
For reprocessing, attach the PROPHYflex perio tip
③
with adapter
②
to the
cleaning adapter PROPHYflex 4 long (
Mat. no. 3.004.6640
)
①
.
▶
Slide the PROPHYflex perio tip into the wrench until it locks into place.
▶
Screw the PROPHYflex perio tip onto the adapter in clockwise direction hold-
ing the wrench with 2 fingers.
▶
Pull the wrench off.
▶
Attach the PROPHYflex perio tip to the cleaning adapter PROPHYflex 4 long
(
Mat. no. 3.004.6640
).
3
2
1
①
PROPHYflex perio tip
②
Adaptor
③
Cleaning adapter
PROPHYflex 4 long
▶
Place the cleaning adapter PROPHYflex 4 long (
Mat. no. 3.004.6640
) on
the receptacle for INTRA connectors.