Instructions for use PROPHYflex 4 Perio Kit - 1.011.9403 Refill PROPHYflex perio tip - 1.010.0287
6 Reprocessing steps in accordance with ISO 17664 | 6.5 Visual inspection and functional check
6.4.2 Automated internal and external cleaning and
internal and external disinfection
KaVo recommends washer disinfectors according to EN ISO 15883-1, which
are operated using alkaline cleaning agents having a maximum pH value of 10.
The validation was conducted with a Miele washer disinfector using the
"VARIO-TD" programme, the "neodisher® mediClean" cleaning agent, the
neodisher® Z" neutralisation agent and the "neodisher® mielclear" rinsing
agent.
▶
For programme settings as well as cleansers and disinfectants to be used,
please refer to the Instructions for Use of the washer disinfector.
6.4.3 Automated Drying
The drying procedure is normally part of the cleaning program of the thermod-
isinfector.
Note
Please observe the instructions for use of the thermodisinfector.
▶
In order to prevent impairment of the KaVo medical device, make sure that
the inside and outside of the device is dry after the end of the cycle.
▶
Remove any residual liquids with KaVo DRYspray.
See also:
2
6.5 Visual inspection and functional check
Note
Plug the adaptor with the PROPHYflex perio tip onto the PROPHYflex with an
empty powder container, and check it.
6.6 Care products and systems - Servicing
NOTICE
Improper care.
Malfunction or property damage.
▶
Do not service the PROPHYflex perio tip and adapter with oils or care spray.
6.7 Packaging
Note
The sterilisation bag must be large enough to accommodate the perio tip
without stretching the bag.
The quality and use of the sterilisation packaging must comply with applic-
able standards and be suitable for the sterilisation procedure!
▶
The medical device must be packed before sterilisation.
19 / 24