Instructions for use SMARTmatic S80 K - 1.011.6771
7 Reprocessing steps in accordance with ISO 17664 | 7.3 Disinfection
▶ Cover the medical device with the KaVo CLEANpac bag, and place it on the cor-
responding care adapter. Press the spray button three times for 2 seconds each
time. Remove the medical device from the spray attachment and let the cleanser
act for one minute.
▶ Afterwards, rinse for 3-5 seconds with KaVo DRYspray.
See also:
2
KaVo CLEANspray / KaVo DRYspray Instructions for Use
Note
KaVo CLEANspray and KaVo DRYspray for manual interior cleaning are only avail-
able in the following countries:
Germany, Austria, Switzerland, Italy, Spain, Portugal, France, Luxembourg, Bel-
gium, Netherlands, United Kingdom, Denmark, Sweden, Finland and Norway.
In other countries interior cleaning can only be carried out with thermodisinfectors in
accordance with EN ISO 15883-1.
7.2.4 Automated internal cleaning
KaVo recommends washer disinfectors in compliance with EN ISO 15883-1 that are
operated with alkaline cleaning agents.
The validations were conducted with the VARIO-TD program, the cleaning agent
neodisher® MediClean and the neutralisation agent neodisher® Z.
▶ For program settings as well as cleansers and disinfectants to be used, please
refer to the Instructions for Use of the thermodisinfector.
▶ In order to prevent damage to the medical device due to residual fluid, make sure
that the inside and outside of the device is dry after the end of the cycle. Remove
any residual liquids from the interior and exterior of the medical device using com-
pressed air.
▶ Immediately after drying, lubricate the medical device with care agents from the
KaVo care system.
The drying procedure is normally part of the cleaning program of the thermodisin-
fector.
Note
Please observe the instructions for use of the thermodisinfector.
7.3 Disinfection
WARNING
Incomplete disinfection.
Infection hazard.
▶ Principally, KaVo recommends carrying out an final disinfection of the unpack-
aged item to guarantee the complete disinfection.
NOTICE
Using the disinfectant bath or chlorine-containing disinfectants.
Malfunction and material damage.
▶ Do not disinfect the device in the disinfection bath or with chlorine-containing dis-
infectants.
15 / 24