background image

Instructions for use SURGmatic S11 L - 1.009.1010 SURGmatic S11 C - 1.009.1005

7 Reprocessing steps in accordance with EN ISO 17664 | 7.6 Packaging

Note

Close the chuck service mode.

Option 1: Place the dental handpieces in the QUATTROcare PLUS 2124 A,
close the front door and start theservice procedure.
Option 2: After three minutes with no service procedure running, the device
automatically switches back to normal service mode.

See also:

2

Servicing with KaVo QUATTROcare PLUS

7.5.3 Servicing with KaVo QUATTROcare

Note

QUATTROcare 2104 / 2104 A is no longer included in the current de-
livery programme.

Follow-up product:

QUATTROcare PLUS 2124 A

Servicing and cleaning device with expansion pressure for the interior cleaning
of inorganic residues and optimum care.
(no validated cleaning of the interior according to German Robert Koch Institute
(RKI) requirements)
KaVo recommends servicing the product as part of the reprocessing after each
use, i.e. after each cleaning, disinfection, and before each sterilisation.

Remove the tool from the medical device.

Service the product in the QUATTROcare.

Care of chucking system

KaVo recommends cleaning and servicing the chuck system once a week.

Remove the tool from the medical device.

Plug the spray nipple of the chuck servicing set onto the QUATTROcare plus
Spray.

Position the tip of the spray nipple in the opening, and apply the spray.

Press the spray key for 1 to 2 seconds.

7.6 Packaging

Note

The sterile goods package must be large enough for the product so that the
packaging is not stretched.
The quality and use of the packaging of the items to be sterilised must satisfy
the applicable standards and be appropriate for the sterilisation process!

The medical device must be packed before sterilisation.

19 / 24

Summary of Contents for SURGmatic S11 L

Page 1: ...Instructions for use SURGmatic S11 L 1 009 1010 SURGmatic S11 C 1 009 1005 ...

Page 2: ...istributed by KaVo Dental Technologies LLC 11727 Fruehauf Drive Charlotte NC 28273 USA Phone 847 550 6800 Fax 847 550 6825 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com ...

Page 3: ...r malfunctions 14 6 2 Troubleshooting 14 6 2 1 Cleaning the spray tube 14 7 Reprocessing steps in accordance with EN ISO 17664 15 7 1 Preparation at the site of use 15 7 2 Non fixing preliminary cleaning of the spray tube 15 7 3 Manual reprocessing 16 7 3 1 Manual internal and external cleaning and internal and external disinfection 16 7 3 2 Manual drying 16 7 4 Automated reprocessing 16 7 4 1 Aut...

Page 4: ...lease contact KaVo Repair Service KaVo Dental Technologies LLC 11727 Fruehauf Drive Charlotte NC 28273 USA Toll free Direct Customer Service 1 888 ASK KAVO 888 275 5286 Email techservice kavokerr com www kavo com Target group The instructions for use are intended for medical professionals in particular dentists and office personnel The section on Commissioning is also intended for service technici...

Page 5: ...ormity GOST R certification Transportation and storage conditions Temperature range Transportation and storage conditions Air pressure Transportation and storage conditions Humidity Protect from moisture Protect from impact HIBC Code Hazard levels The warning and safety notes in this document must be observed to prevent personal injury and material damage The warning notes are designated as shown ...

Page 6: ...RGmatic S11 L 1 009 1010 SURGmatic S11 C 1 009 1005 1 User instructions 6 24 CAUTION In cases which if not prevented could lead to minor or moderate injury NOTICE In cases which if not prevented could lead to material damage ...

Page 7: ...ately before disposal 2 2 Technical condition A damaged device or components could injure patients users and third parties Only operate devices or components if they are undamaged on the outside Check that the device is working properly and is in satisfactory condition before each use Have parts with sites of breakage or surface changes checked by the Ser vice If the following defects occur stop w...

Page 8: ...n of rotation must be present The medical device may only be combined with a treatment centre released by KaVo Comply with the Instructions for Use of the treatment centre 2 4 Qualification of personnel Application of the product by users without the appropriate medical training could injure the patients the users or third parties Make sure that the user has read and understood the instructions fo...

Page 9: ...acial surgery A medical device according to relevant national statutory regulations Proper use According to these regulations this product may only be used for the described application by a properly trained user You need to comply with the following the applicable health and safety regulations the applicable accident prevention regulations these Instructions for use According to these regulations...

Page 10: ...ce can be mounted on all INTRAmatic LUX motors and motors with a connection in accordance with EN ISO 3964 3 3 Transportation and storage conditions Do not store in a refrigerated environment Temperature 20 C to 70 C 4 F to 158 F Relative humidity 5 RH to 95 RH absence of condensation Air pressure 700 hPa to 1060 hPa 10 psi to 15 psi Protect from moisture ...

Page 11: ...king the amount of water CAUTION Overheating of the tooth due to insufficient amount of cooling water Thermal damage to the dental pulp Adjust the water amount for the spray cooling to a minimum of 50 ml min CAUTION Hazard from insufficient amount of spray water Insufficient spray water can cause the medical device to overheat and damage the tooth Check the spray water channels and clean the spray...

Page 12: ... contra angle handpiece while the drive motor is rotating Lightly spray O rings on motor coupling with KaVo Spray Place the medical device on the motor clutch and lock it into place With the SURGmatic S11 L the latch must lock into place audibly Pull on the medical device to make sure that it is securely affixed to the coupling 5 2 Remove the medical device Unlock the medical device from the motor...

Page 13: ...ve the bits to prevent injury and infection Rotate the clamping ring in the direction of the arrow to the backstop and insert the handpiece cutter or polisher into the chuck Turn the clamping ring back into its initial position Check that the cutter or grinder is securely attached by pulling on it 5 4 Remove the handpiece or contra angle drill bit WARNING Hazard from rotating cutter or grinder Lac...

Page 14: ...edical device When the speed drops or is uneven Service the medical device An O ring is missing on the motor coupling Replace O ring See also 2 Instructions for use of motor 6 2 Troubleshooting 6 2 1 Cleaning the spray tube CAUTION Hazard from insufficient amount of spray water Insufficient spray water can cause the medical device to overheat and damage the tooth Check spray water channels and if ...

Page 15: ...ported to reprocessing Do not place in solutions or similar substances 7 2 Non fixing preliminary cleaning of the spray tube Accessories required Demineralised water 30 C 2 C 86 F 3 6 F Nozzle pin Brush e g medium hard toothbrush Disposable syringe Check the patency of the spray tube and clean it using the nozzle needle Mat no 0 410 0931 Rinse the spray tube with at least 20 ml demineralised water...

Page 16: ...e with care agents from the KaVo care system 7 4 Automated reprocessing WARNING Incomplete disinfection Infection hazard Only use disinfection procedures that are verified to be bactericidal fungi cidal and virucidal If the disinfectants used do not meet these requirements the process must be concluded by disinfection of the unit s without packaging using a steam steriliser WARNING Sharp tool in t...

Page 17: ...tions for use of the washer disinfector In order to prevent impairment of the KaVo medical device make sure that the inside and outside of the device is dry after the end of the cycle Immediately after drying lubricate the KaVo medical device with care agents from the KaVo care system 7 5 Care products and systems Servicing WARNING Sharp dental bur or diamond grinder in the medical device Risk of ...

Page 18: ...infection and before each sterilisation Remove the tool from the medical device Service the product in the QUATTROcare PLUS See also 2 Instructions for use KaVo QUATTROcare PLUS Servicing the clamping chuck KaVo recommends servicing the chuck system once a week using the chuck servicing program integrated in the device Note Handpieces must be taken off the service couplings before the chuck servic...

Page 19: ...leaning of the interior according to German Robert Koch Institute RKI requirements KaVo recommends servicing the product as part of the reprocessing after each use i e after each cleaning disinfection and before each sterilisation Remove the tool from the medical device Service the product in the QUATTROcare Care of chucking system KaVo recommends cleaning and servicing the chuck system once a wee...

Page 20: ...emove the product from the steam steriliser after the steril isation cycle The KaVo medical device has a maximum temperature resistance up to 138 280 4 F Select a suitable procedure depending on the available autoclave from the fol lowing sterilisation processes Steriliser with triple pre vacuum at least 3 minutes at 135 C 275 F Drying time 16 min Steriliser using the gravity method at least 10 mi...

Page 21: ...63 Nozzle pin 0 410 0931 Coupling piece 0 593 0361 Spray head INTRA KaVo Spray 0 411 9911 Service coupling for heads QUATTRO care 0 411 7941 Surgery service coupling 1 009 9489 Only for the USA Material summary Mat no KaVo Spray USA and Canada 2113 A 0 411 9660 QUATTROcare plus Spray USA and Canada 2141 P 1 005 4524 Chuck servicing set 1 003 1253 Canada only Material summary Mat No KaVo Spray Cana...

Page 22: ...for defects and their consequences that have arisen or may arise from natural wear improper handling cleaning or maintenance non compliance with operating maintenance or connection instructions calcination or corrosion contaminated air or water supplies or chemical or electrical factors deemed abnormal or impermissible in accordance with KaVo s instructions for use or other manufacturer s instruct...

Page 23: ......

Page 24: ...1 011 1910 Fk 20190311 04 en ...

Reviews: