10
ATTENTION: Follow additional mainte-
nance recommendations to make the service
life of the rechargeable batteries longer.
Maintenance of KaWe PICCOLIGHT:
Proper func-
tioning is provided by KaWe PICCOLIGHT handles with
2 batteries (“АА”-Mignon 1.5 V). To switch the device
on/off, use a sliding switch. To switch the device on
using a sliding switch, move it towards the
dermatoscope’s head. To switch the device off, move
the sliding switch in reverse direction, to reference po-
sition. To replace the battery and lamp, unscrew the
head of KaWe PICCOLIGHT.
Time of application:
The devices are designed for
periodic use, maximum operation time totals 1 minu-
te. There shall be 10-minute pauses between the ap-
plications.
Lamp replacement of KaWe PICCOLIGHT:
For
proper light functioning and quality, use only original
KaWe lamps. Unscrew the head of the dermatoscope
from the handle and remove the lamp. Clean the glass
bulb of the new lamp with alcohol-containing cloth.
The glass bulb shall be clean and free from fingerprints
(grease spots). Put the lamp all the way into its groove
in the side pin. Then screw the investigation head back
to the handle.
Lamp replacement of KaWe EUROLIGHT:
For
proper light functioning and quality, use only original
KaWe lamps. Unlatch the head of the dermatoscope
from the handle and remove the lamp. Now repeat the
steps above in the section on replacing KaWe PICCO-
LIGHT lamp. The glass bulb shall be clean and free from
fingerprints (grease spots). Put the lamp all the way in.
Handling:
Put special moisturizing KaWe Derma-Gel
dermatoscope gel on the skin area of interest. Turn the
light on and set the contact lens on skin without exces-
sive pressing. Even lighting will cover the skin area
with large examination diameter (25 mm). Place the
examined area in the center of the contact lens, if pos-
sible. Bring the eyepiece as close to the eye as possible.
With a free hand, turn the focus ring until the image of
the area of interest becomes sharp. Try looking through
the eyepiece without pressure on eye muscles (not
accommodating them for sharper view). Scales con-
tact lens is designed to control the pigmented skin
changes. To examine raised or palpable skin areas (da-
mages), dermatoscope can be used without the con-
tact lens. For a sharper view you can raise the dermato-
scope over the area of interest but not further than 5
mm.
Technical characteristics:
• Magnifier glass with 10x magnification
• Contact lens diameter: 25 mm
• Focusing / sharpness adjustment: -6...+3.5 diopter
• Halogen lamp aperture: 2.5 V
(KaWe EUROLIGHT D30)
• Vacuum lamp aperture: 2.5 V
(KaWe PICCOLIGHT D)
• LED lamp aperture: 3.5 V
(KaWe EUROLIGHT D30 LED)
Additional directions, maintenance and sto-
rage:
The device will serve for years with proper
handling and storage.