Guarantee:
When used under normal circumstances
and with attention to these instructions we guarantee
this product for two years after the date of purchase
(with the exception of lamps and rechargeable bat-
teries). Should you need further information or should
your instrument require repair, please contact your
dealer.
Warning instructions:
Never bring the der-
matoscope in contact with liquid and make sure that
no liquid penetrates the housing! Since the dermato-
scope is not designed for operations, simple cleaning
and/or disinfection with an alcohol-based surface dis-
infectant is sufficient. Do not use any abrasive cleans-
ing agents! The dermatoscope cannot be sterilized.
Limitation of reprocessing:
Frequent reprocessing
influences the instruments only minimally. The work-
ing life of the instruments usually depends on the wear
and tear and damage by use.
Instructions
Workstation:
Remove dirt accumulation on surfaces
with a disposable/paper cloth.
Storage and transport:
No special requirements
Cleaning preparation:
No special requirements
Manual cleaning:
The exterior surface of the instru-
ment can be cleaned with a damp, soft and lint-free
cloth. Never allow the dermatoscope to come into
contact with fluids and ensure that liquids do not
penetrate the housing. As the dermatoscope is not
intended for use during operational procedures, sim-
ple cleaning and or disinfection with an alcohol-based
surface disinfecting agent is sufficient. Do not use
abrasive cleaning agents! The unit cannot be sterilised.
The magnifying glass is to only be cleaned with glass
cleaner.
Automatic cleaning:
No automatic cleaning pos-
sible/required.
Disinfection:
No special requirements. The disinfec-
tant can be used in accordance with the instructions
given on the label.
Sterilisation:
The measuring scale disc must be
sterilised in the autoclave at a temperature of 134°C
for a minimum of 3 minutes.
Additional information:
If many instru ments are
being sterilised in the autoclave at the same time, the
maximum load given by the manufacturer must not
be exceeded.
Maintenance:
Defective parts must be removed
and/or replaced. If the instrument does not function
properly, follow these steps:
1.
Replace batteries and
ensure that they are charged.
2.
Replace the lamp and
ensure that it works properly.
3.
Send the unit in for
repair.
Check and performance test:
Check all movable
parts such as the frames of the magnifying glasses
and plugin connections to ensure that they all move
smoothly. Turn on the unit. For all units: Visually check
all parts for damage and wear. Keep all electrical con-
tact points clean.
Packaging:
Individually packed in zippered or cloth
bags.