14

PICCOLIGHT

1. Insert batteries (2x Mignon or AA 1.5 V batteries) into 

the handle. Check the polarity! Also follow instructions 

in Section 8: Maintenance and Service.

2. Screw the head onto the handle.

3. Use the sliding switch to turn the lamp on or off.

4.You can change batteries/rechargeable batteries by 

unscrewing the head from the handle. Also follow 

instructions in Section 8: Maintenance and Service.

Duration of operation

The devices are not intended for continuous operation, the ma-

ximum duration of one use is 1 minute with a 10 minutes pause 

between uses.

5.  Preparation for use

1. Check the assembled otoscope for damage, impurities and 

compatibility. Check that the moving parts (i.e. magnifying 

glass frame, plug junctions) move smoothly.

2. Check that the lamp is well secured and functions properly.
3. Check the battery charge.

If all those checks have been successful, you can now use the 

otoscope to examine a patient after sanitizing it and attaching 

a new ear funnel.

6.  Sanitary preparation

Disposable ear funnels are single-use only and
therefore are not to be cleaned or re-used!

Clean reusable ear funnels immediately after use to prevent 

impurities from drying up.

Manual / automated cleaning

Protect the device from moisture!

Never allow the otoscope to come into contact 

with fluids and ensure that liquids do not 

penetrate the housing.

  
The exterior surface of the device can be cleaned with a damp, 

soft and lint-free cloth. Simple cleaning and or disinfection with 

an alcohol-based surface disinfecting agent is sufficient.
•  Do not sterilize the device!
•  Do not use abrasive cleaning agents!
•  The magnifying glass is only to be cleaned with glass cleaner!

No automated cleaning possible/required!

Disinfection

The otoscope can be disinfected by wiping it with a cloth 

dampened with alcohol (e.g. isopropyl). Avoid moisture from 

getting on the device!
Disinfect reusable ear funnels with a suitable disinfectant 

(i.e. Helipur®). Follow the concentrations and reaction times 

indicated by disinfectant manufacturer.

Sterilization

Only reusable ear funnels are to be sterilized!

Proper steam sterilization can only be done after sufficient 

pre-cleaning!

Maximum hot steam sterilization temperature: ...............134 °C
Minimum hot steam sterilization time:  .............................3 Min.

O

N

Summary of Contents for EUROLIGHT

Page 1: ... KaWe PICCOLIGHT Gebrauchsanweisung Otoskope User s Manual Otoscopes Instruction d utilisation Otoscopes Istruzioni per l uso Otoscopi Manual de instrucciones Otoscopios Manual de usuário Otoscópios Инструкция по применению Отоскопы ...

Page 2: ...Sanitary preparation 14 Manual automated cleaning 14 Disinfection 14 Sterilization 14 7 Storage and transport 15 Ambient temperature 15 Relative air humidity 15 8 Service and maintenance 15 Battery change 15 EUROLIGHT COMBILIGHT 15 PICCOLIGHT 15 Lamp change 15 EUROLIGHT F O COMBILIGHT F O PICCOLIGHT F O 15 PICCOLIGHT C 16 EUROLIGHT C COMBILIGHT C 16 9 Accessories and spare parts 16 10 Disposal 17 ...

Page 3: ...ation power etc are not specifically covered herein because there are no notable diffe rences in their servicing 1 Symbols used in this User s Manual and on the product package Safety instructions and warnings Item No Manufacturer s number lot number Manufacturer Please follow this User s Manual Storeindryconditions Single useonly CE conformity mark GOST R certification of goods imported into Russ...

Page 4: ...ly and to be disposed of immediately after use 4 Assembly and service EUROLIGHT COMBILIGHT 1 Insert the battery or the rechargeable battery into the handle Check for correct polarity and for correct lamps 2 5 V or 3 5 V Also follow the instructions of Section 8 Maintenance and Service 2 C10 options Push the head all the way down hold it in place and tighten the screw Check the screw groove on the ...

Page 5: ...els are single use only and therefore are not to be cleaned or re used Clean reusable ear funnels immediately after use to prevent impurities from drying up Manual automated cleaning Protect the device from moisture Never allow the otoscope to come into contact with fluids and ensure that liquids do not penetrate the housing The exterior surface of the device can be cleaned with a damp softandlint...

Page 6: ...ltage caused by wear leads to diminishing illumination and in some cases to blinking LED lamps In both cases replacing the batteriesisrecommended Usehighqualityalkalinebatteriesor fully charged rechargeable batteries to replace the old ones Please follow the User s Manual for the rechargeable batteries or charger devices EUROLIGHT COMBILIGHT a battery 2x Baby C type batteries b rechargeable batter...

Page 7: ...e black plastic bushing from the lamp carefully 3 Clean the bulb of the replacement lamp with alcohol The glass bulb should be clean and free of fingerprints grease spots 4 Screw a replacement lamp into the head and cover it with plastic bushing 9 Accessories and spare parts Disposable ear funnels anthracite Ø 2 5 mm 4 0 mm black Ø 2 5 mm 4 0 mm Reusable ear funnels black Ø 2 0 mm 2 5 mm 3 0 mm 3 ...

Page 8: ...lectronic devicesandrechargeablebatteriesmustbetrans ferred to a designated waste collection point for recycling User and patient instructions Please inform the manufacturer and the competent authorities of the relevant member state where the user and or patient is located immediately about any serious product related incidents 11 Annex Warranty The warranty for LED lamps is five years after the s...