16
6.2 Adjusting the arm
• adjust the clamping lever
If the light body is hard to move or does not stay in the desired posi-
tion, the braking action of the clamping lever must be adjusted.
7. TECHNICAL DATA
7.1 Photometric data
Central illumination intensity
at a distance of 0.5 meters
60,000 Lux
Central illumination intensity
at a distance of 1 meter
14,000 Lux
Light spot diameter d
10
at a distance of 0.5 m
103 mm
Light spot diameter d
10
at a distance of 1 m
228 mm
Light spot diameter d
50
at a distance of 0.5 m
60 mm
Light spot diameter d
50
at a distance of 1 m
103 mm
Residual illumination intensity with one shutter
0 %
Residual illumination intensity with two shutters
78 %
Residual illumination intensity at the bottom of a
normed tube
100%
Residual illumination intensity at the bottom of a
normed tube and one shutter
0 %
Residual illumination intensity at the bottom of a
normed tube and two shutters
78 %
Illumination depth 60 %
1180 mm
Colour rendering index CRI
95
Colour rendering index R
9
94
Max. irradiation intensity in the field at a distance
of 1.0 m
51 W/m²
Max. irradiation intensity in the field at a distance
of 0.3 m
330 W/m²
Light spot size
10 cm
Colour temperature (Kelvin)
4000 K
Heat build-up near the head
0.5 °C
Number of LEDs
7
Working lifespan of the LEDs
40,000 h
Diameter of the light body
22 cm
Note:
The technical data are subject to certain deviations. Due
to production–related reasons, the actual values can vary
slightly from the above-named values. The values for R
a
can
deviate by approx. ± 5%. The colour temperature values can
deviate by approx. ± 200K.
7.2 Electrical data
Power input
10 W
Operating voltage
24 V DC
Amperage
0,42 A
7.3 Electrical installation information
The KaWe examination light is exposed to a current peak when it is
switched on and is therefore equipped with a power adapter.
Warning! This light is a Protection Class I device. To avoid the
risk of electric shock, this unit must be connected to a ground-
ed electrical outlet.
7.4 Weights
Examination Lamp
Weight
MASTERLIGHT LED 2000
6.0 kg*
* with 5-foot stand
7.5 Environmental conditions
Operation
Min.
Max.
Temperature
+10° C
+30° C*
Relative humidity
30 %
75 %
Air pressure
700 hPa
1060 hPa
* for higher temperatures please contact us