17
Loosen the safety screw on the upper
side of the handle. Remove the
KaWe
PICCOLIGHT
®
F.O.
– otoscope head
from the handle and pull out the lamp.
If necessary, clean the glass bulb of the
new lamp with alcohol. The bulb
should be clean and free of finger prints
and oil or grease. The new lamp must
be inserted all the way.
Loosen the safety screw on the upper
side of the handle. Unscrew the
KaWe
PICCOLIGHT
® – ophthalmoscope
head from the handle. Remove the used lamp and reinsert a new one,
ensuring that it is pushed in all the way.
Technical Specifications for the KaWe MedCenter® 5000
Admissible ambient temperature ...........................+10 °C to +40 °C
Relative humidity ..................................................................10 to 75%
Air pressure .................................................................700 to 1060 hPa
Protection type...............................................................................IP X0
Protection class ....................................................................................II
Power plug-AC adapter
.................Type UE15WCP1-045200SPA
Wide range input voltage ................................100-240 V~, 50/60 Hz
Output voltage ......................................................................... 4.5 V DC
Output current ........................................................................ max. 2 A
Cord length ......................................................................................3 m
Base module (with frame)
Dimensions of the base module ...................204.5 x 96.5 x 135.5 mm
Spiral power cord length .............. 2800 mm (700 mm coiled length)
Weight .....................................................................................ca. 810 g
Extension module
Dimensions of the additional module...............80 x 96.5 x 135.5 mm
Spiral power cord length ................2800 mm (700 mm coiled length
Weight .....................................................................................ca. 370 g
Ear funnel module
Dimensions of the ear funnel module...............80 x 96.5 x 135.5 mm
Weight .....................................................................................ca. 200 g
IEC 60601-1-1
IEC 60601-1-2
Manufacturer:
KaWe
Warning:
This device is maintenance free. For this reason, the housing can only
be opened on the side end of the module’s frame for the addition of
add-on parts. Repairs are only to be carried out by a qualified expert.
Never use more than two handles for each KaWe MedCenter® 5000
at the same time! The unit does not have any replacement parts.
If damage is apparent, disconnect the power cord immediately. The
adapter serves in this case as the separator and must be disconnected
from the mains supply if damage is apparent. The charging station
is not designed for use in environments in which explosions may
occur, in the presence of strong magnetic fields, or in surgery. The
spiral cable can cause people to trip. Do not place it where people
may be walking. Only KaWe-brand components are to be used.
Modifications to this device are not allowed.
While performing examinations, the user is not to touch the
conducting parts such as the electric plug, screw closure on the
diagnostic unit, etc.
When the light is left on for longer than 1 minute, the metal on the
upper part of the handle becomes very hot and when touched can
hurt or cause burns. If the handle left on surfaces that are temperature
sensitive, there may be a risk of fire. Ensure that after each use, the
handle is returned to its holder and the light is turned off. After using
for an extended period of time, allow enough time for the handle to
cool.
Instructions
Storage and Transport:
Please store this device in its packaging
prior to assembly in order to avoid contact with impurities or dust.
Cleaning Preparation:
No specific requirements.
Manual Cleaning:
The external parts of the charging station and
the diagnostic equipment can be cleaned with a damp, soft, lint-free
cloth. Soapy water or other non-aggressive agents may be used.
Automatic Cleaning:
No automated cleaning possible / necessary.
Disinfection:
The housing of the unit may be wiped with a surface
disinfectant. Always wipe the surface dry.
Maintenance:
Remove and replace damaged parts.
Inspection and Verification of Proper Operation:
Inspect the
movable parts (such as the magnifier frame and the plug connection)
to ensure that they move easily. Turn the diagnostic device on.
All devices: perform a visual inspection for damage and wear.