23
18. Technical Information
Classification
Applied part: type BF
Protection classification
IP 41
Dimensions
W 110 x H 84 x L 120 mm
Nominal output current
Adjustable 1 ..... 25 mA
Load resistance
max. 1.5 kΩ
Power input
1.5 W
Power supply
Li-ion rechargeable battery
Operation mode
Continuous operation
Weight
Charging plug: approx. 0.10 kg
iontophoresis device: approx. 0.36
kg
Safety features
Electronic current limiting device
for the treatment current, Current is
cut off when the treatment circuit
is broken, System interrupt for
when the treatment circuit is not
closed
Current type
Continual current or pulsed current
up to a max. of 25 mA
Ambient conditions
Operating temperature
+ 10 °C to + 40 °C
Storage temperature
0 °C to + 40 °C
Relative humidity
30 % to 75 %
Air pressure
700 hPa to 1060 hPa
19. Guarantee
We guarantee this product (except for the Li-ion rechargeable battery)
for 48 months. The prerequisite for this guarantee is a proper use of the
product in accordance with this instruction manual. The guarantee is
valid from date of purchase. The bill of sale serves as proof in this case.
Wear parts - such as electrodes, foam pads and sponge pockets - are
not covered by the warranty!
Before returning the instrument in the event of a malfunc-
tion, the possible causes described in the section
“Troubleshoot ing” should all be examined thoroughly.
Should we discover upon inspecting the returned device, a
reason for the malfunction that is described in the user’s
manual that could have been easily solved by the customer
themself, we will charge an inspection fee. (This fee in-
cludes the cost of labour, postage and packing, as well as
VAT.)
EMC instructions: Medical electrical equipment is subject to
special safety measures so as to ensure safe operation with
respect to electromagnetic compatibility. The following in-
formation helps to ensure safe operation.
20. Guidance and manufacturer’s declaration – electromagnetic emmisions
The SwiSto3 is designed to be used in the types of electromagnetic environments listed below. The customer or user is responsible for
ensuring that this device is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – Guidance
RF emissions
CISPR 11
Group 1
The SwiSto3 device uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B
The SwiSto3 device is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/flicker
emissions IEC 61000-3-3
Complies
Summary of Contents for SwiSto 3
Page 27: ...27 ...