www.kawemed.com
21
Charge only with the KaWe REF 05.19170.002 charging plug!
Charge the battery completely before using for the first time.
The charging time for a completely discharged battery is
approx. 10 hours. When charging, the battery symbol blinks
and the charging symbol on the display appears in the middle
of the screen. The device cannot be used while it is being
charged. The charging progress can be seen on the battery
symbol.
To perform a treatment, place the SwiSto3 iontophoresis
device on a flat and dry surface.
Place the water trays in an appropriate location.
Place the electrodes (REF 05.19040.021) in the water trays.
Place the foam pads on the electrodes such that the electrodes
are completely covered.
Use the connector cables to connect the electrodes with
the output jacks (+ and -) on the side of the device. When
connecting the cables to the electrodes, they should be
plugged in with a simultaneous twisting motion.
Fill both of the tubs with a minimum of 3-4 cm of lukewarm
tap water. Take special care that the trays do not overflow or
have the potential to overflow when the hands or feet are put
into the water.
For the underarm perspiration treatment, insert the plate
electrodes into the sponge pockets and soak them well in
water.
Power-on the SwiSto3 iontophoresis device by pressing the
push-button switch. Hold the button down for approx. 1
second until the display illuminates. Current treatment values
will be shown on the display.
The SET button is used to select each of the individual settings.
Each of the blinking values can be changed with the up and
down buttons. The following can be changed: the current
intensity in steps of 1 mA, continuous current / pulsing current
and the treatment time in min:sec.
If the display is not readable or the unit appears to be
damaged in any way, send the device in to the manufacturer
immediately.
9. Treatment of foot, hand and
underarm perspiration
Prepare device for use as described in Section 8 “Start-up/
preparation”.
Power the device on by pressing the power button for one
second.
- Treatment of foot and hand perspiration: In each of the tap
water-filled trays, place the palm of a hand or the sole of
a foot onto the foam pad-covered plate electrode. Directly
touching the electrodes is not dangerous; in order to avoid
spots of excessive current density, it is recommended
however, that as much of the surface of the body part
to be treated as possible rests on the electrodes and the
pressure is uniformly distributed.
- Treatment of underarm perspiration: Wet the sponge
pockets evenly with tap water! In order to avoid excessively
high current densities, care should be taken to ensure
that the contact area is as large as possible and that the
pressure is evenly distributed. Insert the plate electrodes
(REF 05.19080.001) into the sponge pockets.
The parameters that can be selected will begin to blink on the
display. The current intensity for the treatment can be selected
using the UP and DOWN buttons. --> Confirm selection by
pressing the SET button. --> Select pulsing current with the
UP button and continuous current with the DOWN button.
--> Confirm selection by pressing the SET button. --> Set
the treatment time in minutes (MIN) using the UP and DOWN
buttons. --> Confirm selection by pressing the SET button.
--> Set the treatment time in seconds (SEC) using the UP
and DOWN buttons. --> Confirm selection by pressing the
SET button. --> Setting selection is completed when the
treatment starts, the time countdown begins and the clock
symbol begins to blink.
Beginning the treatment
For foot or hand perspiration: The treatment begins by closing
the circuit by immersing the hands or feet in the water tubs.
For underarm perspiration: The treatment for underarm
perspiration begins by closing the circuit by inserting the
electrodes (encased in the sponge pockets) under the armpits.
1) Raise the upper arm and insert the electrode into the arm
pit.
2) The electrode is then held in place by the lowered upper
arm. Insure before you lower the upper arm that the
electrode and connecter cable are completely inserted into
the sponge pocket, as direct contact between metal and
skin can lead to localised burns. Especially when treating
underarm perspiration, please keep the pressure created
by the upper arm as constant as possible for the entire
duration of the treatment, which can be up to 30 minutes.
ENGLISH
Summary of Contents for SwiSto3
Page 3: ...Deutsch English Fran ais Italiano Espa ol Portugu s 6 17 18 29 30 41 42 53 54 65 66 77 78 89...
Page 4: ......
Page 79: ...www kawemed com 79 BF DIN IEC 601 1 VDE 0750 1 LOT IP41 1 0 REF SN 7 2 KaWe 1...
Page 80: ...www kawemed com 80 0 2 KaWe 4 8 800...
Page 86: ...www kawemed com 86 SwiSto3 CISPR 11 1 SwiSto3 CISPR 11 B SwiSto3 IEC 61000 3 2 A IEC 61000 3 3...
Page 90: ......
Page 91: ......