www.kawemed.com
24
Important information!
Parts such as electrodes, foam pads, sponge pockets and
connector cables are subject to a certain amount of wear and
may require replacement after prolonged use. Check all parts
for damage before each application, especially the connecting
lines. If corroded, the cables are to be replaced. If the contact
plugs on the SwiSto3 become corroded, send the device in
for inspection.
The integrated rechargeable battery may only be replaced
by the manufacturer. Should the battery be defective (will
not charge, discharges too quickly), send the device in for
inspection/repair.
Warning:
Modifying the SwiSto3 medical electrical device is
not allowed! In order avoid the risk of electric shock, only the
included charging plug (REF 05.19170.002) may be used to
charge the unit.
14. Storage while not in use
In order to prevent the SwiSto3 iontophoresis device from
becoming damaged, it should be stored along with its
accessories in its original packaging if not in use for a
prolonged period of time.
15. Transport and storage
When transported, the SwiSto3 iontophoresis device is to be
packaged such that it is protected from damage. If the device
is not to be used for a prolonged period of time, it should
be charged completely before being stored. The device and
accessories are to be stored in a dry location between 0 °C
and +40 °C.
16. Disposal
The non-accessible part of the SwiSto3 iontophoresis device
contains electrical and electronic components. For this reason,
at the end of its service life, the device must be brought to
an appropriate waste disposal site in accordance with local
regulations or returned to the manufacturer.
17. Technical specifications
Classification
Applied part: type BF
Protection classification
IP 41
Dimensions
W 110 x H 84 x L 120 mm
Nominal output current
Adjustable 1 ..... 25 mA
Load resistance
max. 1.5 kΩ
Power input
1.5 W
Power supply
Li-ion rechargeable battery
Operation mode
Continuous operation
Weight
Charging plug: approx.
0.10 kg
iontophoresis device: approx.
0.36 kg
Safety features
Electronic current limiting
device for the treatment
current, Current is cut off
when the treatment circuit is
broken, System interrupt for
when the treatment circuit is
not closed
Current type
Continual current or pulsed
current up to a max. of 25 mA
Operating temperature
+ 10 °C to + 40 °C
Storage temperature
0 °C to + 40 °C
Relative humidity
30 % to 75 %
Air pressure
700 hPa to 1060 hPa
ENGLISH
Summary of Contents for SwiSto3
Page 3: ...Deutsch English Fran ais Italiano Espa ol Portugu s 6 17 18 29 30 41 42 53 54 65 66 77 78 89...
Page 4: ......
Page 79: ...www kawemed com 79 BF DIN IEC 601 1 VDE 0750 1 LOT IP41 1 0 REF SN 7 2 KaWe 1...
Page 80: ...www kawemed com 80 0 2 KaWe 4 8 800...
Page 86: ...www kawemed com 86 SwiSto3 CISPR 11 1 SwiSto3 CISPR 11 B SwiSto3 IEC 61000 3 2 A IEC 61000 3 3...
Page 90: ......
Page 91: ......