Guidance and manufacturer’s declaration – electromagnetic emmisions
The SwiSto3 is designed to be used in the types of electromagnetic environments listed below. The customer or user is
responsible for ensuring that this device is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – Guidance
RF emissions
CISPR 11
Group 1
The SwiSto3 device uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The SwiSto3 device is suitable for use in all establishments,
including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/flicker
emissions IEC 61000-3-3
Complies