Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The SwiSto3 is intended for use in the electromagnetic environment specified below.
The customer or user is to assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment –
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact discharge
± 8 kV air discharge
± 6 kV contact discharge
± 8 kV air discharge
Floors should be wood, concrete,
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
±2 kV for power supply
lines
±1 kV for input/output
lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1kV voltage
line to line
± 2kV voltage line
to ground
± 1kV voltage
line to line
± 2kV voltage line
to ground
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short
interruptions, and
voltage variations on
power-supply input
lines
IEC 61000-4-11
<5 % UT
for 1/2 cycle
(>95 % dip)
40 % UT
for 5 cycles
(60 % dip)
70 % UT
for 25 cycles
(30 % dip)
<5 % UT for
5 seconds
(>95 % dip)
<5 % UT
for 1/2 cycle
(>95 % dip)
40 % UT
for 5 cycles
(60 % dip)
70 % UT
for 25 cycles
(30 % dip)
<5 % UT for
5 seconds
(>95 % dip)
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the device
requires continued OPERATION during
a power mains interruption, it is
recommended that the device be
powered by an uninterruptible power
supply or battery.
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m
3 A/m
If interference occurs, it may be
necessary to place the SwiSto3
further away from the sources of
power frequency magnetic fields or to
install a magnetic shield: The power
frequency magnetic field should be
measured at the designated site to
ensure that it is not large enough to
cause interference.
Note: UT is the ac mains voltage prior to application of the test level.
ENGLISH
Summary of Contents for SwiSto3
Page 3: ...Deutsch English Fran ais Italiano Espa ol Portugu s 6 17 18 29 30 41 42 53 54 65 66 77 78 89...
Page 4: ......
Page 79: ...www kawemed com 79 BF DIN IEC 601 1 VDE 0750 1 LOT IP41 1 0 REF SN 7 2 KaWe 1...
Page 80: ...www kawemed com 80 0 2 KaWe 4 8 800...
Page 86: ...www kawemed com 86 SwiSto3 CISPR 11 1 SwiSto3 CISPR 11 B SwiSto3 IEC 61000 3 2 A IEC 61000 3 3...
Page 90: ......
Page 91: ......