www.kawemed.com
29
Recommended separation distances between portable and mobile RF communications equipment and the SwiSto3
The SwiSto3 is intended for use in the electromagnetic environment described below. The customer or the user of the
SwiSto3 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the SwiSto3 as recommended below, according to the maximum output power
of the communications equipment.
Separation distance (meters) according to frequency of transmitter
(m)
Rated maximum output power of
transmitter
(W)
150 kHz to 80 MHz
d = 1.2 √P
80 MHz to 800 MHz
d = 1.2 √P
800 MHz to 2.5 GHz
d = 2.3 √P
0.01
0.12
0.12
0.23
0.10
0.38
0.38
0.73
1
1.20
1.20
2.30
10
3.80
3.80
7.30
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m)
can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (w) according to the transmitter manufacturer.
Note 1:
For the calculation of the separation distance for transmitters with a frequency range of 80 MHz to 2.5 GHz, an
additional factor
of 10/3 is used in order to decrease the likelihood that mobile/portable communications equipment could cause interference if
it is inadvertently brought into patient areas.
Note 2:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
ENGLISH
Summary of Contents for SwiSto3
Page 3: ...Deutsch English Fran ais Italiano Espa ol Portugu s 6 17 18 29 30 41 42 53 54 65 66 77 78 89...
Page 4: ......
Page 79: ...www kawemed com 79 BF DIN IEC 601 1 VDE 0750 1 LOT IP41 1 0 REF SN 7 2 KaWe 1...
Page 80: ...www kawemed com 80 0 2 KaWe 4 8 800...
Page 86: ...www kawemed com 86 SwiSto3 CISPR 11 1 SwiSto3 CISPR 11 B SwiSto3 IEC 61000 3 2 A IEC 61000 3 3...
Page 90: ......
Page 91: ......