Chapter 7 Starting Standalone Manual Calibration
Kaye HTR-420 User Manual
35
Display Range:
The following information is shown in the display range:
Current External Control
Reference Probe value.
Current test point / set point.
Probable remaining test
duration for the current test
point.
Step counter for stepped
operation.
Reference Status.
Reference used (ext).
The test continues running until its desired parameters have been achieved then
automatically stops and ramps back to its default startup temperature.
Note:
If data logging
is enabled, a screen appears prompting the user to enter the test temperature target value.
A
Test Task Completed Successfully
window appears.
To stop testing manually, touch and slide the left arrow next to
Stop
all the way to the left.
To obtain logged data, tap on the data icon to bring up the
Select Log Data
screen. Tap on
the desired file then the check mark in the lower right-hand corner.
The
View Log Data
window appears providing user with the following four options:
o
Diagram
o
Table
o
Export Log Data
o
Export All Log Data
Choose the desired option(s).
Following completion of test, power the unit down.
Note:
If the temperature value is greater than 40°C then a message appears stating, “calibrator is
reaching safe temperature – please wait.” Once the value drops below 40°C then a new message
appears informing that “you can now turn off the calibrator.”
After the testing process:
Allow the calibrator to cool to room temperature.
Note the instructions in the section Test Tasks on page 29.
Clean the calibrator.