Chapter 10: Sensor Verification
Kaye ValProbe RT User Manual
96
Figure 108:
Verification Study Screens
Note
:
It is important to ensure live data is displayed from the Logger sensors and IRTD
before starting a Verification. This guarantees a good RF connection between Loggers
and base station and so you can be sure that each logger receives the Start
Verification command.
If using a Dry Block reference, place the logger sensors into the Dry Block and then proceed
to next step to Start. If using an oil bath press
Start
and then after sensor data is displayed,
fully submerse Loggers into oil bath. In an oil bath, it is possible that live data from the Logger
is no longer received. The verification study will not be affected, as the data is stored in the
Loggers and can be read out after the study is finished.
Press
Start
on the Temperature Verification screen to start a Verification study.
When you start a Verification, the Kaye ValProbe RT software detects whether a Kaye
temperature reference and an IRTD are connected. Both are required for a fully automatic
operation. If there is no IRTD connected it is not possible to run a Verification study. If the
Kaye temperature reference is not found, the system does not allow automated Verification. In
this case, only a manual verification is possible. If neither can be found, Verification process
cannot be run.
Figure 109: Auto / Manual Verification Prompt
If Auto calibration is selected the Base Station sends the corresponding temperature setpoint
to the Temperature Reference
Summary of Contents for ValProbe RT
Page 1: ...Validation Kaye ValProbe RT User Manual MXXX EN Rev 1 0 May 2020 ...
Page 2: ... No content intended for this page ...
Page 4: ... No content intended for this page ...
Page 12: ......
Page 163: ...Kaye ValProbe RT User Manual 150 ...
Page 166: ... No content intended for this page ...