28592_EDEK
ZA0044-0224-1_EN
34
5.7
Binary outputs
Binary outputs (BO)
4 binary outputs
3 x closer
1 x changeover
Contact specification (EN60947-4-1, -5-1)
Configuration
3 x SPST (Single Pole Single Throw)
1 x SPDT (Single Pole Double Throw)
Nominal voltage
250 VAC
Nominal current
6 A
Nominal load AC1
1500 VA
Nominal load AC15, 230 VAC
300 VA
Interrupting power
DC1, 30/110/220 V
6/0,2/0,12 A
Number of switching operations AC1
≥ 60·10
3
electrical
Electrical Isolation
Isolated from all internal potentials
Input resistance
> 100 kΩ
Electrical safety IEC61010
300 V CAT II
5.8
Temperature Input
Temperature input PT100 / PT1000 / KTY – (from firmware version 2.2
Contacting measurement sensor
(software setting)
2 wire
3 wire
4 wire
Update rate
1 Second / 1 Hz
Resolution
15 Bit
Burden
1,9 kΩ
Accuracy
0.05% FSR
Technical data