background image

   
 

             

SHENZHEN KEDIDA ELECTRONICS CO., LTD

 

CT-6020/CT-6020L   

Pen pH Meter Instruction Manual

  

 Caution

 

 

Please read the manual carefully before using the meter. 

 

The glass electrode at the bottom of the pen is fragile, please use it carefully after taking off the protection cap. Any 

damage may cause the invalid of the electrode. 

 

The duration period of the glass electrode is one year from the date of purchase whether you use it or not, please 

change the electrode in time. 

 

The electrode can not be used under the dry condition, always soak the electrode into the distilled water or purified 

water for10 – 30 minutes to activate before use.   

 

The meter is ready to be used after the model shows on the screen. 

FEATURES

 

* Range:      0.00 to 14.00 pH 

* LCD indicates both pH & Temp. 

* Pen type digital pH meter, all in one, pH electrode is 
included, easy to use 

* Data hold function for freezing the desired value 

* Water proof and protection 

* Auto power off without any operations after 10 mins 

*  Supplied  with  a  set  of  pH  4.01,  pH  7.00,  pH  10.01 
buffers 

*Microcomputer  circuit,  intelligent  function,  high 
accuracy 

* Easy to change the pH electrode 

* Compact size, light weight 

* Options temperature measurement, 

  or 

 

* Power supply by DC 1.5 V battery (LR44, BAT ) x 4 PCs 

*  Build  in  temperature sensor,  ATC  (auto  temperature 
compensation) 

*Available  for  wide  applications,  such  as  aquarium, 
beverage, 

fish 

hatcheries, 

food 

processing, 

photography,  laboratory,  quality  control,  school  & 
colleges, swimming pools, water conditions 

* Auto calibration for pH 4.01, pH 7.00 and pH 10.01   

 

* Intelligent backlight for CT-6020L– Sensing technology automatically turn on or turn off LED screen backlighting to 

react to ambient light levels. 

SPECIFICATIONS 

Display 

LCD, size : 20 mm x 27 mm. Consumption 

Measurement    Range 

pH

 

0.00 to 14.00 pH

 

Temp. 

0 to 50 

℃ 

( 32 to 122 

℉ 

Resolution

 

pH

 

0.01 pH

 

Temp.

 

0.1 

,    1 

 

Accuracy

 

pH

 

±  0.02 pH                     

* After calibration

 

Temp.

 

±  1 

, ±  2 

 