7.5  2-phase models 

 

The OAF models can be made for use on 2 phases of a 3-phase plus neutral supply. If a diagram is 
required please contact Carbolite. 

 

8.0  F

USES 

& P

OWER 

S

ETTINGS

 

8.1 

Fuses 

 

F1-F3: Refer to the circuit diagrams

F1

 

Internal 
supply 
fuses 

Fitted if supply cable fitted. 
Fitted on board to some types of EMC filter. 

on-board and up to 16 Amps:  
32mm x 6mm type F 
other: GEC Safeclip 

F2

 

Auxiliary 
circuit 
fuses 

Fitted on board to some types of EMC filter.  
May be omitted up to 25Amp/phase supply 
rating. 

2 Amps glass type F 
On board: 20mm x 5mm  
Other: 32mm x 6mm  

F3

 

Heat Light 
fuses 

May be omitted up to 25 Amp/phase supply 
rating. 

2 Amps glass type F  
32mm x 6mm  

 

Customer 
fuses 

Required if no supply cable fitted. 
Recommended if cable fitted. 

See rating label for amperage; 
see table below for fuse rating. 

 

 

Model Phases  Volts 

Supply 

Fuse 

rating 

AAF 11/3 

1-phase 

200-240 

10A 

AAF 11/7 
and 
OAF 11/1 

1-phase 
1-phase 
2-phase+N 

220-240 
110-120 
380/220-415/240 

20A 
40A 
10A per phase 

AAF 11/18 

1-phase 

220-240 

30A 

 1-phase 

200-208 

40A 

 

3-phase+N 

380/220-415/240 

12.5A per phase 

 

3-phase delta 

220-240 

20A per phase 

 

3-phase delta 

200-208 

25A per phase 

OAF 11/2 

1-phase 
2-phase+N 

220-240 
380/220-415/240 

20A 
10A per phase 

GSM 11/8 

1-phase 
1-phase 

220-240 
110-120 

12.5A 
25A 

VMF 10/6 

1-phase 

220-240 

12.5A 

1-phase

 

8.2 

Power Settings 

The settings of the power limit parameter 

OP.Hi

 are as follows. 

model           volts:  200 

208 

220, 280  230, 400  240, 415  254, 440 

AAF 

11/3 

100 

98 87 80 73 65 

AAF 

11/18 

100 100 100 100 94  83 

all 

other 

models 

100 100 100 100 100 89 

 

18 

MF06 

Summary of Contents for CARBOLITE AAF 11/18

Page 1: ...d to the electricity supply CONTENTS Section page 1 0 Descriptions 2 2 0 Installation 3 3 0 Operation 4 4 0 Maintenance 7 5 0 Repairs Replacements 8 6 0 Fault Analysis 14 7 0 Circuit Diagrams 15 8 0 F...

Page 2: ...r flow through the chamber 1 3 GSM 11 7 This is a conventional muffle furnace with a Silica chamber This gives a cleaner environment than a standard ceramic muffle 1 4 VMF 10 6 The Volatile Matter Fur...

Page 3: ...models and AAF 11 7 may alternatively be ordered for use on two phases out of a 3 phase with neutral supply The model AAF 11 18 may also be ordered for 3 phase supply see section 2 4 Check the furnace...

Page 4: ...perature controller see the controller manual OAF 11 2 the slave controller should be set to an offset temperature of zero see 3 2 Overtemperature option only If the overtemperature controller has not...

Page 5: ...AAF OAF GSM VMF 3 3 MF06 5...

Page 6: ...used to heat materials which emit smoke or fumes regularly heat it up to maximum temperature for one hour without load to burn away the soot Materials such as case hardening compounds and other react...

Page 7: ...respondence please quote the serial number model type and voltage given on the rating label of the furnace The serial number and model type are also given on the front of this booklet when supplied wi...

Page 8: ...Guidance Note EH46 We can provide further information on request Alternatively our service division can quote for any repairs to be carried out at your premises or ours 5 3 Temperature Controller Rep...

Page 9: ...AAF OAF GSM VMF 5 5 MF06 9...

Page 10: ...section above which refers to the thermocouple colour coding Unfasten the electrical connections to the inner chamber brick box and remove the thermocouple and sheath see thermocouple replacement ins...

Page 11: ...ermocouple cables from their terminal blocks Straighten all the element tails Note that these become brittle with use Unscrew the screws fixing the metal back piece from the inner chamber and withdraw...

Page 12: ...at the furnace is controlling properly to rule out the possibility that the previous element failed because of a fault elsewhere in the control circuit 5 9 VMF 10 6 Element Replacement See section 5 2...

Page 13: ...s good emission of some fumes may be expected 5 11 Fuse Replacement Fuses are marked on the circuit diagram section 7 0 with type codes e g F1 F2 A list of the correct fuses is given in section 8 1 De...

Page 14: ...o lights glowing on the controller The SUPPLY light is ON The controller may be faulty or not receiving a supply due to a faulty switch or a wiring fault The SUPPLY light is OFF Check the supply fuses...

Page 15: ...d PE L N F2 F1 SSR F3 F3 safety switches A heat on e l e m e n t s for AAF 11 18 see diagram 7 4 7 2 OAF 11 2 single phase 7 3 Higher Voltages e g 254V For single phase models of 254V or above a contr...

Page 16: ...AAF OAF GSM VMF 7 4 16 MF06...

Page 17: ...n t s e l e m e n t s N L L N L N thermocouple Fuses or EMC filters Supply Terminals Circuit no 3 Circuit no 2 N L N L N L3 L2 L1 N PE L Circuit no 1 3 phase neutral A model made to this diagram can...

Page 18: ...Customer fuses Required if no supply cable fitted Recommended if cable fitted See rating label for amperage see table below for fuse rating Model Phases Volts Supply Fuse rating AAF 11 3 1 phase 200 2...

Page 19: ...l 41 41 541 5877 Fax 41 41 541 5878 Email General Information Sales info kesn ch Accounts Department accounts kesn ch USA O ce 233 Rogue River Highway 425 Grants Pass Oregon 97527 5429 USA Tel 1 503 2...

Reviews: