Page 35
DOC0945 (22 August 2022 15:28)
www.keit.co.uk
Back to Table of Contents
Inspection of the sealing weld will require removal of the flange and spectrometer from the
pressure vessel and can be completed as part of regular inspection/maintenance of the
spectrometer and pipework.
Do not use impact (e.g. a hammer) to separate components if the gasket becomes stuck to
the flange and hub.
The lap-joint flange may be manufactured from Hastelloy or stainless steel. Care should be
exercised to ensure that it is not corroded by aggressive chemicals during maintenance or
installation.
5.7. Inclined Stand - ASM0916
The inclined stand is designed for using the spectrometer as a bench top instrument for offline
sampling. Angled at 35°, the probe can be positioned into beakers or other vessels for ease
of use. It can also be configured for two heights to give added flexibility in how it is used.
Figure 31: IRmadillo on an inclined stand
5.7.1. Safety
WARNING
The spectrometer and inclined stand assembly weighs ~24kg. Use
caution when handling to avoid injury.
5.7.2. Installation
Assemble the inclined stand to the desired height; two positions are possible. For technical
drawings, contact Keit at enquiries@keit.co.uk.