(H) Mounting holes
The spectrometer must be supported using the four M8 mounting holes in its base. Optional
feet are provided.
(I) Power cable gland
This is the mains power connection to the spectrometer. More detail on how to connect power
to the spectrometer can be found in Section 2.1 Powering the IRmadillo.
4.1.1. Dust caps
The communication flying leads are provided with dust caps to protect the RJ45 connectors.
When the cable is not in use, these should be attached to reduce the risk of potential damage
to the connectors that may be caused by the local environment.
4.2. Controller specifications
Table 3 - Specifications of the IRmadillo integrated controller
Category
Specification
Integrated
controller
The integrated controller has been optimised for the rugged operation of
the IRmadillo spectrometer; it is not a general-purpose personal computer.
Keit does not recommend the installation of additional software; changes
to the configuration may affect performance and reliability.
User
accounts
There are two user accounts, one with system administrator privileges and
a second ‘standard’ user. The standard user account starts automatically
and is the default account for the KeitSpec software application. The system
administrator account is provided to make changes to the operating system
configuration and install additional software, if required.
Windows
updates
The controller is running Windows 10 LTSC with automatic updates
disabled by default. The local administrator should choose when/if to
update the operating system. A strategy will be required to manage
updates, particularly if the controller is connected to a network.
Anti-virus
There is no anti-virus installed. The controller should only be connected to
internal, protected networks. Install your preferred anti-virus if required
4.3. Storage conditions
When installed with purge and power, the IRmadillo can thermally control its internal
temperatures up to the limit specified on the product label. However, if the system is to be
stored before its use Keit recommends an air-conditioned environment. The environment must
be non-condensing, with a relative humidity below 95%. The IRmadillo or storage container
should also be out of direct sunlight.
WARNING
Storing the IRmadillo unpowered and unpurged in unsuitable
conditions can cause damage to the instrument. If in doubt,
please consult with Keit