9
Use
▲
or
▼
to scroll through selections, ENTER to
select, or ESC to cancel selection.
TRIGGER MODE: select multiple (continu-
ous with one trigger) or one-shot (one read-
ing per trigger).
TRIGGER SOURCE: select external,
manual, disabled, or IEEE-488 X, GET or
talk triggers.
TRIGGER DELAY: program desired trig-
ger delay (time between trigger and read-
ing in one-shot trigger mode): 0sec-
999.999sec.
TRIGGER INTERVAL: program desired
trigger interval (time between individual
readings in multiple trigger mode): 0.01
sec-999.999sec.
BUFFER
Use STORE to start/stop buffer data storage. Store
size of 1-1024 selects linear buffer (storage stops when
size limitation is reached); store size of 0 selects cir-
cular buffer (buffer wraps around and overwrites old
data after 1024 readings are stored). Unit stores read-
ings at programmed trigger interval in multiple trigger
mode or one reading per trigger in one-shot trigger
mode.
Press RECALL to access stored buffer readings and
buffer statistics. Readings include buffer location and
time stamp. Statistics include number of samples,
buffer average, maximum, minimum, and standard
deviation.