Operation
3-3
Scanner card detection
Power-up detection
The scanner card is detected only at power-on. If the card is plugged in after the power is
turned on, the multimeter will not recognize the presence of the card.
CAUTION
Plugging in the scanner card with power turned on may result in damage to
both the Model 2000-SCAN-20 and the multimeter.
If the card is not present at power-on, scanner bus commands or queries will generate a
“Missing hardware error,” and front panel operations pertaining to the scanner will inform you
that no scanner is present.
Scanner option bus query
*OPT? is an IEEE 488.2 common query that determines whether or not a scanner card is
installed. Also, the Model 2000-20 multimeter can distinguish between a Model-2000-SCAN
10-channel card and the Model-2000-SCAN-20 20-channel card. The response to this query is
as follows:
•
0: Scanner not installed.
•
0,10CH: Model 2000-SCAN 10-channel card installed.
•
0,20CH: Model 2000-SCAN-20 20-channel card installed.
Refer to the multimeter manual for more details on using the *OPT? query.
Summary of Contents for 2000-SCAN-20
Page 11: ......
Page 12: ...1 General Information General Information ...
Page 17: ...1 6 General Information ...
Page 18: ...2 CardConnections andInstallation CardConnec tionsand Installation ...
Page 35: ...2 18 Card Connections and Installation ...
Page 36: ...3 Operation Operation ...
Page 51: ...3 16 Operation ...
Page 52: ...4 Service Information Service Information ...
Page 63: ...4 12 Service Information ...
Page 64: ...5 Replaceable Parts Replace ableParts ...
Page 67: ...5 4 Replaceable Parts ...
Page 68: ......
Page 69: ......
Page 70: ......
Page 71: ......
Page 72: ......
Page 73: ......
Page 74: ...A Specifications Section A Specs 7 8 03 4 32 PM Page 1 ...
Page 79: ......
Page 80: ......