Operation
3-5
Opening channels
Front panel
Use the front panel OPEN key to open all the card channels.
Remote interface
Use the :ROUTe:OPEN:ALL command to open all the channels. The command is as follows:
PRINT #1, "output 16; :route:open:all"
Use the :ROUTe:MULTiple:OPEN command to open specified channels. For example, if you
want to open channels 1, 3, and 5, use the following command:
PRINT #1, "output 16; :route:multiple:open (@ 1,3,5)"
NOTE
Opening channel 21 selects the 2-pole operating mode.
Selecting measurement function
Use the following procedure to select a specific measurement function for an internal
channel:
1.
Use the
and
keys to select the desired channel.
2.
Select the desired measurement function with the keys on the front panel. The various
measurement functions available are:
DCV: DC volts
PERIOD: period
ACV: AC volts
TEMP: simulated temperature
Ω
2W: 2-wire ohms
CONT: continuity
Ω
4W: 4-wire ohms
: diode test
(channels 1 through 10)
FREQ: frequency
Summary of Contents for 2000-SCAN-20
Page 11: ......
Page 12: ...1 General Information General Information ...
Page 17: ...1 6 General Information ...
Page 18: ...2 CardConnections andInstallation CardConnec tionsand Installation ...
Page 35: ...2 18 Card Connections and Installation ...
Page 36: ...3 Operation Operation ...
Page 51: ...3 16 Operation ...
Page 52: ...4 Service Information Service Information ...
Page 63: ...4 12 Service Information ...
Page 64: ...5 Replaceable Parts Replace ableParts ...
Page 67: ...5 4 Replaceable Parts ...
Page 68: ......
Page 69: ......
Page 70: ......
Page 71: ......
Page 72: ......
Page 73: ......
Page 74: ...A Specifications Section A Specs 7 8 03 4 32 PM Page 1 ...
Page 79: ......
Page 80: ......