3-10
Operation
4-wire resistance tests
More precise measurements over a wider range of system and DUT conditions can be
obtained using the 4-wire measurement scheme shown in
from the Model 2000-20 Multimeter are routed through the scanner to the resistor under test.
The extra set of sense leads minimizes the effects of voltage drops across the test leads, greatly
reducing measurement errors with lower DUT resistances. Note, however, that an extra two
poles of switching are required for each resistor tested. For this reason, only ten resistors can be
tested using this configuration.
Figure 3-2
4-wire resistance test connections
DMM 2000-20
Rear Panel
1
DUTs
(10)
R
A. Test Configuration
B. Simplified Equivalent Circuit
10
11
20
1
10
HI
LO
Sense LO
HI
Sense HI
LO
Sense LO
2000-20
2000-SCAN-20
DUT
2000-SCAN-20
Out A
Out B
Sense HI
Ω
Summary of Contents for 2000-SCAN-20
Page 11: ......
Page 12: ...1 General Information General Information ...
Page 17: ...1 6 General Information ...
Page 18: ...2 CardConnections andInstallation CardConnec tionsand Installation ...
Page 35: ...2 18 Card Connections and Installation ...
Page 36: ...3 Operation Operation ...
Page 51: ...3 16 Operation ...
Page 52: ...4 Service Information Service Information ...
Page 63: ...4 12 Service Information ...
Page 64: ...5 Replaceable Parts Replace ableParts ...
Page 67: ...5 4 Replaceable Parts ...
Page 68: ......
Page 69: ......
Page 70: ......
Page 71: ......
Page 72: ......
Page 73: ......
Page 74: ...A Specifications Section A Specs 7 8 03 4 32 PM Page 1 ...
Page 79: ......
Page 80: ......