Operation
3-13
Figure 3-5
Voltage attenuation by path isolation resistance
Magnetic fields
When a conductor cuts through magnetic lines of force, a very small current is generated.
This phenomenon will frequently cause unwanted signals to occur in the test leads of a scanning
system. If the conductor has sufficient length, even weak magnetic fields like those of the earth
can create sufficient signals to affect low-level measurements. Two ways to reduce these effects
are: (1) reduce the lengths of the test leads, and (2) minimize the exposed circuit area. In extreme
cases, magnetic shielding may be required. Special metal with high permeability at low flux den-
sities (such as mu metal) is effective at reducing these effects.
Even when the conductor is stationary, magnetically-induced signals may still be a problem.
Fields can be produced by various signals such as the AC power line voltage. Large inductors
such as power transformers can generate substantial magnetic fields, so care must be taken to
keep the switching and measuring circuits a good distance away from these potential noise
sources. At high current levels, even a single conductor can generate significant fields. These
effects can be minimized by using twisted pairs, which will cancel out most of the resulting
fields.
Electromagnetic interference (EMI)
The electromagnetic interference characteristics of the Model 2000-20 Multimeter comply
with the electromagnetic compatibility (EMC) requirements of the European Union (EU)
directives as denoted by the CE mark. However, it is still possible for sensitive measurements to
be affected by external sources. In these instances, special precautions may be required in the
test setup.
Sources of EMI include:
•
Radio and television broadcast transmitters.
•
Communications transmitters, including cellular phones and handheld radios.
•
Devices incorporating microprocessors and high-speed digital circuits.
•
Impulse sources as in the case of arcing in high-voltage environments.
R
DUT
E
DUT
R
PATH
E
OUT
R
PATH
R
DUT
R
PATH
+
=
E
DUT
E
OUT
Summary of Contents for 2000-SCAN-20
Page 11: ......
Page 12: ...1 General Information General Information ...
Page 17: ...1 6 General Information ...
Page 18: ...2 CardConnections andInstallation CardConnec tionsand Installation ...
Page 35: ...2 18 Card Connections and Installation ...
Page 36: ...3 Operation Operation ...
Page 51: ...3 16 Operation ...
Page 52: ...4 Service Information Service Information ...
Page 63: ...4 12 Service Information ...
Page 64: ...5 Replaceable Parts Replace ableParts ...
Page 67: ...5 4 Replaceable Parts ...
Page 68: ......
Page 69: ......
Page 70: ......
Page 71: ......
Page 72: ......
Page 73: ......
Page 74: ...A Specifications Section A Specs 7 8 03 4 32 PM Page 1 ...
Page 79: ......
Page 80: ......