4-4
Service Information
Recommended equipment
summarizes the equipment necessary for performance verification and the
application for each item.
Path resistance tests
Perform the following steps to verify that the path resistance of each channel is within the
specified values.
NOTE
To perform this test, you must custom-wire a cable to connect to the scanner card
connections.
1.
Connect the HI and LO inputs of all 20 channels together to make one common input
using #22AWG solid copper jumper wires (
).
NOTE
Make sure all jumper wires are clean and free of oxides.
2.
Connect the output cables to OUT A HI and LO as shown in
.
3.
With the power off, install the scanner card in a Model 2000-20 Multimeter.
Table 4-1
Recommended verification equipment
Description
Model/part
Specifications
Application
Multimeter
Electrometer with voltage
source
Sensitive Digital Voltmeter
Triax cable (unterminated)
Low-thermal cable
#22AWG solid copper wire
#22AWG Teflon-insulated
stranded wire
Triax adapter (3-slot male to
2-lug female)
Keithley 2000-20
Keithley 6517A
Keithley 182
Keithley 7025
Keithley 1484
—
—
Keithley 6171
100
Ω
; 100ppm of
reading; 40ppm of
range
20pA, 200pA; 1.0%
100V source; 0.15%
3mV; 60ppm
—
—
—
—
—
Control card, measure path resistance
Path isolation
Contact potential
Path isolation
Contact potential
Path resistance
Path isolation
Path isolation
Summary of Contents for 2000-SCAN-20
Page 11: ......
Page 12: ...1 General Information General Information ...
Page 17: ...1 6 General Information ...
Page 18: ...2 CardConnections andInstallation CardConnec tionsand Installation ...
Page 35: ...2 18 Card Connections and Installation ...
Page 36: ...3 Operation Operation ...
Page 51: ...3 16 Operation ...
Page 52: ...4 Service Information Service Information ...
Page 63: ...4 12 Service Information ...
Page 64: ...5 Replaceable Parts Replace ableParts ...
Page 67: ...5 4 Replaceable Parts ...
Page 68: ......
Page 69: ......
Page 70: ......
Page 71: ......
Page 72: ......
Page 73: ......
Page 74: ...A Specifications Section A Specs 7 8 03 4 32 PM Page 1 ...
Page 79: ......
Page 80: ......