2-14
Card Connections and Installation
Card installation and removal
This paragraph explains how to install and remove the Model 2000-SCAN-20 card assembly
from the Model 2000-20 Multimeter.
WARNING
Installation or removal of the Model 2000-SCAN-20 should be performed
only by qualified service personnel. Failure to recognize and observe
standard safety precautions could result in personal injury or death.
NOTE
To prevent contamination to the scanner card that could degrade performance,
handle the card assembly only by the card edges.
Scanner card installation
Perform the following steps and refer to
to install the card assembly in the
multimeter:
WARNING
Turn off power to all instrumentation (including the multimeter) and
disconnect all line cords. Make sure all power is removed and any stored
energy in the external circuitry is discharged.
1.
Pry out the two fasteners holding the OPTION SLOT cover plate to the rear panel of the
multimeter. Remove the cover plate.
2.
Slide the card edges, component side up, into the guide rails inside the multimeter.
3.
With the ejector arms in the unlocked position, carefully push the card all the way
forward until the arms engage the ejector clips. Push both arms inward to lock the card
into the multimeter.
4.
After installation, connect the output leads to the multimeter as discussed in the
following paragraph.
Summary of Contents for 2000-SCAN-20
Page 11: ......
Page 12: ...1 General Information General Information ...
Page 17: ...1 6 General Information ...
Page 18: ...2 CardConnections andInstallation CardConnec tionsand Installation ...
Page 35: ...2 18 Card Connections and Installation ...
Page 36: ...3 Operation Operation ...
Page 51: ...3 16 Operation ...
Page 52: ...4 Service Information Service Information ...
Page 63: ...4 12 Service Information ...
Page 64: ...5 Replaceable Parts Replace ableParts ...
Page 67: ...5 4 Replaceable Parts ...
Page 68: ......
Page 69: ......
Page 70: ......
Page 71: ......
Page 72: ......
Page 73: ......
Page 74: ...A Specifications Section A Specs 7 8 03 4 32 PM Page 1 ...
Page 79: ......
Page 80: ......