Operation
3-11
Current measurements
Although the Model 2000-SCAN-20 is not intended for direct current measurements, on-card
shunt resistors can be installed for channels 1 and 11 to allow for indirect current measurements.
shows the typical configuration for current measurements. Refer to
manual for specific information on recommended shunt resistors and their installation.
Figure 3-3
Connections for current measurements
The current through each shunt resistor can be calculated as follows:
I = V/R
Where: I = current being determined
V = voltage measured by multimeter
R = shunt resistance value
Instead of manually calculating current, set up a Model 2000-20 mX + b math calculation
function. (Here, m = 1/R and b = 0). The multimeter will then display the actual current value.
Refer to the multimeter manual for details on using math functions.
DMM 2000-20
Rear Panel
DUTs (2)
V
A. Test Configuration
B. Simplified Equivalent Circuit
2000-20
2000-SCAN-20
DUT
Out A
A
A
A
HI
LO
User-installed
Shunts
User-installed
Shunt
CH 1
CH 11
2000-SCAN-20
Summary of Contents for 2000-SCAN-20
Page 11: ......
Page 12: ...1 General Information General Information ...
Page 17: ...1 6 General Information ...
Page 18: ...2 CardConnections andInstallation CardConnec tionsand Installation ...
Page 35: ...2 18 Card Connections and Installation ...
Page 36: ...3 Operation Operation ...
Page 51: ...3 16 Operation ...
Page 52: ...4 Service Information Service Information ...
Page 63: ...4 12 Service Information ...
Page 64: ...5 Replaceable Parts Replace ableParts ...
Page 67: ...5 4 Replaceable Parts ...
Page 68: ......
Page 69: ......
Page 70: ......
Page 71: ......
Page 72: ......
Page 73: ......
Page 74: ...A Specifications Section A Specs 7 8 03 4 32 PM Page 1 ...
Page 79: ......
Page 80: ......